Z-1084-2015Class II

The MAYFIELD Ultra 360 Base Unit for head positioner, one unit per package

Integra LifeSciences Corporation / initiated 2014-12-19 / Terminated / root cause: manufacturing process

Integra LifeSciences Corporation began a recall of The MAYFIELD Ultra 360 Base Unit for head positioner, one unit per package. Product Usage: This product provides the ability to position and hold a patient... on . FDA classified it as Class II and gives the reason as "Through an investigation of complaints, Integra found that there is the possibility for the locking handle mechanism of the Base Units to fail during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1084-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
The MAYFIELD Ultra 360 Base Unit for head positioner, one unit per package. Product Usage: This product provides the ability to position and hold a patient head while the patient is in the prone, supine, lateral or park-bench and sitting positions. It is intended to be used during diagnostic examinations or surgical procedures where a rigid support between the surgical table and a headrest or skull clamp is necessary and positional freedom is required.
Product code
FWZ Operating Room Accessories Table Tray
Reason for recall
Through an investigation of complaints, Integra found that there is the possibility for the locking handle mechanism of the Base Units to fail during use.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Integra sent an Urgent - Voluntary Medical Device Recall letter dated December 19, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to take the following actions: 1. Review your MAYFIELD equipment and determine if you have any MAYFIELD Ultra 360 Base Units 2. Once identified we recommend that you discontinue use and remove them from service until they are repaired by an authorized Integra Repair Center 3. Complete the attached Recall Acknowledgement and Repair Form 4. Upon receipt of the Recall Acknowledgement and Repair Form, Customer Service will contact you and provide an RMA number and instructions for returning the productor questions call 609-936-6822. For questions contact Customer Service: USA 800-654-2873 Outside USA 609-936-5400
Quantity in commerce
1053 Units
Distribution
Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, and DC. (25 government consignees) and in the foreign countries: CANADA, ALGERIA, ARAB EMIRATES, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CHILE, CHINA, COLUMBIA, CYPRESS, DENMARK, DOMINICAN REPUBLIC, FRANCE, GERMANY, GREAT BRITAN, HONG KONG, HUNGARY, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN , JORDAN, KOREA, LIBYA, MACEDONIA, MAYLASIA, MEXICO, NETHERLANDS, POLAND, PORTUGAL, PUERTO RICO, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI LANKA, TAIWAN, THAILAND, and TURKEY.
Product codes and lots
All Lot numbers of Mayfield Ultra 360 Base Units, Catalogue #'s A2009 and A2029.   Affected Lot Numbers For Catalogue No. A2009  73149, 73355, 73356, 74396, 74730, 74398, 74399, 75796, 75816, 74523, 75838, 75859, 74524, 75873, 74890, 75939, 74892, 76420, 74893, 74894, 74895, 76156, 77225, 76157, 76158, 77831, 76159, 77957, 76160, 78012, 76161, 78548, 77292, 78666, 79067, 79068, 79069, 79071, 80013, 79073, 81782, 79074, 81918, 79075, 79748, 80094, 82194, 80095, 80668, 80669, 84171, 80671, 84288, 82736, 82737, 82738, 85896, 82739, 86400, 82740, 82741, 82742, 84732, 84733, 84734, 84735, 85693, 86817, 86818, 86819, 86820, 86821, 86822, 89453, 89455, 89456, 89457, 92296, 92297, 92298, 92299, 92300, 92301, 94653, 94654, 94655, 99798, 98558, 98559, 98560, 98561, 98562, 94658, 103127, 103128, 103129, 103130, 103131, 104890, 114978, 115809, E19251, 123051  Affected Lot Numbers For Catalogue No. A2029 84180, 84487, 85519, 85815, 85631, 84887, 84888, 86533, 86534, 84889, 86828, 86829, 89302, 86830, 86831, 86832, 86833, 86834, 89534, 89535, 89536, 89537, 89538, and 94167.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1084-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1084-2015Class IITerminatedVoluntary: Firm initiated