Trident PSL HA Solid Back 50 mm Includes Dome Hole Plug Ref 540-11-50E Intended for cementless application in...
Stryker Howmedica Osteonics Corp. began a recall of Trident PSL HA Solid Back 50 mm Includes Dome Hole Plug Ref 540-11-50E Intended for cementless application in a total hip arthroplasty procedure. This system... on . FDA classified it as Class II and gives the reason as "Five customer complaints were received for the same batch of Trident PSL Shells (part number 540-11-50E lot 49073501) alleging that when the packages were opened there were no dome holes plugs present". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1043-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Trident PSL HA Solid Back 50 mm Includes Dome Hole Plug Ref 540-11-50E Intended for cementless application in a total hip arthroplasty procedure. This system is a two piece ( a shell and a liner) design that is assembled during surgery. The shell is press-fit into the patients prepared acetabulum and a linger (polyethylene or alumina ceramic) is then locked into the shell. The femoral head component of the total hip system articulates with the acetabular liner.
- Product code
- MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
- Reason for recall
- Five customer complaints were received for the same batch of Trident PSL Shells (part number 540-11-50E lot 49073501) alleging that when the packages were opened there were no dome holes plugs present.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Stryker sent an Urgent Medical Device letter dated January 7, 2015, with a Recall Notification/Acknowledgement Form via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete and return the attached Product Recall Acknowledgement Form to [email protected] or fax to 1-855-251-3635. Customers should return the affected product to: Attention: Regulatory Compliance Stryker Orthopaedics 325 Corporate Drive Mahwah, New Jersey 07430 Customers with questions should call 201-831-6365. For questions regarding this recall call 201-831-5000.
- Quantity in commerce
- seven units
- Distribution
- Nationwide Distribution including NY, OH, PA, and TX.
- Product codes and lots
- 510K 983382 Lot code 49073501
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1043-2015 | Class II | Terminated | Voluntary: Firm initiated |