Z-1043-2015Class II

Trident PSL HA Solid Back 50 mm Includes Dome Hole Plug Ref 540-11-50E Intended for cementless application in...

Howmedica Osteonics Corp / initiated 2014-12-15 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Trident PSL HA Solid Back 50 mm Includes Dome Hole Plug Ref 540-11-50E Intended for cementless application in a total hip arthroplasty procedure. This system... on . FDA classified it as Class II and gives the reason as "Five customer complaints were received for the same batch of Trident PSL Shells (part number 540-11-50E lot 49073501) alleging that when the packages were opened there were no dome holes plugs present". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1043-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Trident PSL HA Solid Back 50 mm Includes Dome Hole Plug Ref 540-11-50E Intended for cementless application in a total hip arthroplasty procedure. This system is a two piece ( a shell and a liner) design that is assembled during surgery. The shell is press-fit into the patients prepared acetabulum and a linger (polyethylene or alumina ceramic) is then locked into the shell. The femoral head component of the total hip system articulates with the acetabular liner.
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
Five customer complaints were received for the same batch of Trident PSL Shells (part number 540-11-50E lot 49073501) alleging that when the packages were opened there were no dome holes plugs present.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stryker sent an Urgent Medical Device letter dated January 7, 2015, with a Recall Notification/Acknowledgement Form via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete and return the attached Product Recall Acknowledgement Form to [email protected] or fax to 1-855-251-3635. Customers should return the affected product to: Attention: Regulatory Compliance Stryker Orthopaedics 325 Corporate Drive Mahwah, New Jersey 07430 Customers with questions should call 201-831-6365. For questions regarding this recall call 201-831-5000.
Quantity in commerce
seven units
Distribution
Nationwide Distribution including NY, OH, PA, and TX.
Product codes and lots
510K 983382  Lot code 49073501

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1043-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1043-2015Class IITerminatedVoluntary: Firm initiated