Z-1002-2015Class II
Anti-Lambda APC-H7, Catalog Number 656157
Becton, Dickinson and Company, BD Biosciences / initiated 2014-12-19 / Terminated / root cause: supplier
BD Biosciences, Systems & Reagents began a recall of Anti-Lambda APC-H7, Catalog Number 656157. Product Usage: Intended for in vitro flow cytometric immunophenotyping of hematologic disorders on . FDA classified it as Class II and gives the reason as "Two lots of the Anti-Lambda APC-H7 antibody are contaminated with CD38 antibody". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1002-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Product
- Anti-Lambda APC-H7, Catalog Number 656157. Product Usage: Intended for in vitro flow cytometric immunophenotyping of hematologic disorders.
- Product code
- MVU Reagents, Specific, Analyte
- Reason for recall
- Two lots of the Anti-Lambda APC-H7 antibody are contaminated with CD38 antibody.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- BD sent a Recall notification letters dated December, 2014 to affected customers via Certified mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to inspect inventory for affected lots, immediately remove affected reagents from their laboratory, request replacement product and complete and return the enclosed Tracking/Verification form, per the instructions. For questions contact BD Customer Support at 855-236-2772.
- Quantity in commerce
- 15 units
- Distribution
- US Nationwide Distribution in the states of CA, NY, and MI
- Product codes and lots
- Lot numbers 4155632 with expiry date of 2015-06-30. Lot 42889687 was not shipped to customers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1002-2015 | Class II | Terminated | Voluntary: Firm initiated |