da Vinci Si Surgical System, IS3000; - Wall chart for EndoWrist Stapler 45 System. Intended to assist in the...
Intuitive Surgical, Inc. began a recall of da Vinci Si Surgical System, IS3000; - Wall chart for EndoWrist Stapler 45 System. Intended to assist in the accurate control of Intuitive Surgical Endoscopic... on . FDA classified it as Class II and gives the reason as "The Wall chart has been updated because it was noted that Wall Chart (PN 551524-03 Rev A) was inconsistent with the User Manual Adendum (PN 550986-07_B) with regards to neceassary sterilization dry time". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1048-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- da Vinci Si Surgical System, IS3000; - Wall chart for EndoWrist Stapler 45 System. Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- The Wall chart has been updated because it was noted that Wall Chart (PN 551524-03 Rev A) was inconsistent with the User Manual Adendum (PN 550986-07_B) with regards to neceassary sterilization dry time.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Medical Device Recall Replacement letters were sent to all customers on January 15, 2015. The letter identified the reason for the recall, the risk to health, affected products, and actions to be taken by the customer or user and the firm. Customers were asked to discard the existing EndoWrist Stapler Documentation and replace it with the new Documentation provided by the firm. Intuitive Surgical will ship replacement instruments at no charge to customers. Then customers were asked to reunite the replacement instruments with the additional components. The firm will retrain customers on the updated instructions for use. A Clinical Sales Representative or Intuitive Surgical Customer Service at 1-800-876-1310, should be contacted if further information or support is needed concerning the recall notification.
- Quantity in commerce
- 1,034
- Distribution
- Nationwide Distribution
- Product codes and lots
- PN 410298-05, -06, -07, -08, -09