Z-1048-2015Class II

da Vinci Si Surgical System, IS3000; - Wall chart for EndoWrist Stapler 45 System. Intended to assist in the...

Intuitive Surgical, Inc. / initiated 2015-01-15 / Terminated / root cause: labeling and UDI

Intuitive Surgical, Inc. began a recall of da Vinci Si Surgical System, IS3000; - Wall chart for EndoWrist Stapler 45 System. Intended to assist in the accurate control of Intuitive Surgical Endoscopic... on . FDA classified it as Class II and gives the reason as "The Wall chart has been updated because it was noted that Wall Chart (PN 551524-03 Rev A) was inconsistent with the User Manual Adendum (PN 550986-07_B) with regards to neceassary sterilization dry time". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1048-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
da Vinci Si Surgical System, IS3000; - Wall chart for EndoWrist Stapler 45 System. Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
The Wall chart has been updated because it was noted that Wall Chart (PN 551524-03 Rev A) was inconsistent with the User Manual Adendum (PN 550986-07_B) with regards to neceassary sterilization dry time.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Medical Device Recall Replacement letters were sent to all customers on January 15, 2015. The letter identified the reason for the recall, the risk to health, affected products, and actions to be taken by the customer or user and the firm. Customers were asked to discard the existing EndoWrist Stapler Documentation and replace it with the new Documentation provided by the firm. Intuitive Surgical will ship replacement instruments at no charge to customers. Then customers were asked to reunite the replacement instruments with the additional components. The firm will retrain customers on the updated instructions for use. A Clinical Sales Representative or Intuitive Surgical Customer Service at 1-800-876-1310, should be contacted if further information or support is needed concerning the recall notification.
Quantity in commerce
1,034
Distribution
Nationwide Distribution
Product codes and lots
PN 410298-05, -06, -07, -08, -09

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1048-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1048-2015Class IITerminatedVoluntary: Firm initiated