Z-1013-2015Class II

Syngo RT Therapist: The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear...

Siemens Medical Solutions USA, Inc. / initiated 2015-01-02 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo RT Therapist: The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-Ray photon and... on . FDA classified it as Class II and gives the reason as "Combination of CTVision with syngo RT Therapist / syngo RT Oncologist 4.3.SP1 automatic registration in Adaptive Targeting might result in wrong offset calculations. Applying this offset can lead to patient...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1013-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo RT Therapist: The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Combination of CTVision with syngo RT Therapist / syngo RT Oncologist 4.3.SP1 automatic registration in Adaptive Targeting might result in wrong offset calculations. Applying this offset can lead to patient mistreatment. Cone Beam imaging is not affected by this problem.
Root cause tag
software
FDA root cause
Software Design Change
Action
A customer advisory notice, dated January 2, 2015, was sent to end users which identified the product, problem, and action to be taken. It was recommended that customers use only manual registration with the 6 degrees of freedom option disabled, which provides the correct offset values. A software update will be deployed once a solution is available.
Quantity in commerce
3
Distribution
UT, WI, NY
Product codes and lots
model # 08162815, serial #s 10010, 10046, 10586

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1013-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1013-2015Class IITerminatedVoluntary: Firm initiated