Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1255-2015Class II

RAPIDLab 1265 System; Siemens Material Number: 10491395, 10321852, 10335524. The Rapidlab 1200 systems are...

Siemens Healthcare Diagnostics Inc. / initiated 2014-12-30 / Terminated / root cause: software

Siemens Healthcare Diagnostics Inc began a recall of RAPIDLab 1265 System; Siemens Material Number: 10491395, 10321852, 10335524. The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare... on . FDA classified it as Class II and gives the reason as "Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1255-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
RAPIDLab 1265 System; Siemens Material Number: 10491395, 10321852, 10335524. The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb.
Product code
MQM Bilirubin (Total And Unbound) In The Neonate Test System
Reason for recall
Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
Root cause tag
software
FDA root cause
Software design
Action
An Urgent Field Safety Notice was sent by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on December 30, 2014 both in the United States and elsewhere for communication with affected customers. In addition, all affected domestic customers were sent a hard copy of the Urgent Field Safety Notice via Federal Express on December 30, 2014. All affected customers outside the US will be provided a copy of an Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. This notice informs consignees of the issue. Consignees are also asked to complete a field effectiveness check form and return it Siemens via fax (312) 275-7795.
Quantity in commerce
2863 devices
Distribution
Worldwide Distribution - US nationwide including Puerto Rico, Canada, Mexico, Albania, Algeria, Argentina, Austria, Bahrain, Belgium, Bosnia Herzegovina, Botswana, Brazil, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iraq, Ireland, Israel, Italy, Ivory Coast, Kazakhstan, Kuwait, Latvia, Lebanon, Lesotho, Lithuania, Luxembourg, Macedonia, Netherlands, New Caledonia, Norway, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, U.A.E., United Kingdom, Vatican City, Venezuela, White Russia/Belarus.
Product codes and lots
All serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1255-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1252-2015Class IITerminatedVoluntary: Firm initiated
Z-1253-2015Class IITerminatedVoluntary: Firm initiated
Z-1254-2015Class IITerminatedVoluntary: Firm initiated
Z-1255-2015Class IITerminatedVoluntary: Firm initiated