Z-1042-2015Class II

Zimmer PERSONA Tibial Articular Surface Inserter; Item # 42-5299-001-00. Manual surgical instrument for knee...

Zimmer GmbH / initiated 2014-11-25 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of Zimmer PERSONA Tibial Articular Surface Inserter; Item # 42-5299-001-00. Manual surgical instrument for knee arthroplasty (orthopedic) on . FDA classified it as Class II and gives the reason as "Potential for fracture of the tip of the product. The fractured tip is approximately 4mm x 7mm x 3.5mm in size. Retained tip may result in: pain, soft tissue irritation, soft tissue damage, or increased wear due to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1042-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer PERSONA Tibial Articular Surface Inserter; Item # 42-5299-001-00. Manual surgical instrument for knee arthroplasty (orthopedic).
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Potential for fracture of the tip of the product. The fractured tip is approximately 4mm x 7mm x 3.5mm in size. Retained tip may result in: pain, soft tissue irritation, soft tissue damage, or increased wear due to contact with the fractured tip; Revision surgery to remove the fractured tip
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 11/25/2014, URGENT MEDICAL DEVICE RECALL LOT SPECIFIC notifications were sent to the affected distributors, hospital Risk Managers and Surgeons with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers, Surgeons, and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.
Quantity in commerce
1,091
Distribution
Worldwide distribution: US nationwide, including Puerto Rico, South Korea, Singapore, Dominican Republic, Canada, Japan, and Germany.
Product codes and lots
Item No. 42-5299-001-00 Affected Lots: 56572679; 56573170; 56574318; 56575056; 56572918; 56573378; 56574538; 56575057; 56572975; 56573379; 56574540; 56575058; 56572978; 56574123; 56574618; 56575059; 56572981; 56574124; 56574619; 56575060; 56572982; 56574223; 56574620; 56575061; 56573071; 56574291; 56574629; 56575062; 56573115; 56574292; 56574630; 56575065; 56573116; 56574316; 56574631; 56575080; 56573154; 56574317; 56575018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1042-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1042-2015Class IITerminatedVoluntary: Firm initiated