510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1031-2015Class II

HomeChoice/HomeChoice Pro Automated Personal Cycler, for Automated Peritoneal Dialysis. Product codes 5C8310...

Baxter Healthcare Corporation / initiated 2014-10-21 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of HomeChoice/HomeChoice Pro Automated Personal Cycler, for Automated Peritoneal Dialysis. Product codes 5C8310, 5C8310R, R5C8320, and 5C4474 on . FDA classified it as Class II and gives the reason as "The keypad buttons on HomeChoice devices may be activated without the operator pressing them". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1031-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
HomeChoice/HomeChoice Pro Automated Personal Cycler, for Automated Peritoneal Dialysis. Product codes 5C8310, 5C8310R, R5C8320, and 5C4474.
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
The keypad buttons on HomeChoice devices may be activated without the operator pressing them.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An Urgent Device Correction notification was sent to consignees on October 31, 2014. Within the notification, Baxter Healthcare asked customers to do the following: 1) Contact Baxter Technical Services to arrange for the exchange (swap) of any affected HomeChoice PRO Automated PD System devices. 2) Complete the attached customer reply form and return it to Baxter by either fax or scanned e-mail. For general questions regarding this communication, contact The Center for One Baxter at 900-422-9837, Monday through Friday between the hours of 8:00 AM and 5:00 PM Central Time.
Quantity in commerce
200
Distribution
AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI and District of Columbia
Product codes and lots
all codes distributed between 4/15/2014 and 6/11/2014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1031-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1031-2015Class IITerminatedVoluntary: Firm initiated