Z-1058-2015Class II

Puritan Bennett 980 Ventilator System, Model No. PB980 Ventilator (980xxxxxxxx), The Puritan BennettTM 980...

Medtronic, Inc. / initiated 2015-01-12 / Terminated / root cause: software

Covidien LP (formerly Nellcor Puritan Bennett Inc.) began a recall of Puritan Bennett 980 Ventilator System, Model No. PB980 Ventilator (980xxxxxxxx), The Puritan BennettTM 980 Ventilator System is designed for use on Neonatal... on . FDA classified it as Class II and gives the reason as "Covidien is issuing a voluntary field action for all Puritan Bennett 980 ventilators due to occasional GUI transient resets that last approximately 30 seconds". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1058-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Puritan Bennett 980 Ventilator System, Model No. PB980 Ventilator (980xxxxxxxx), The Puritan BennettTM 980 Ventilator System is designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in a hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilator support, delivered invasively or noninvasively, to patients who require the following types of ventilator support: "Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs) "Assist/ Control, SIMV or Spontaneous modes of ventilation.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Covidien is issuing a voluntary field action for all Puritan Bennett 980 ventilators due to occasional GUI transient resets that last approximately 30 seconds.
Root cause tag
software
FDA root cause
Software design
Action
Covidien sent an Urgent Field Safety Notice on January 12, 2015, to all affected customers. Customers were requested to forward the recall letter to anyone they may have transferred or assigned for use one of the potentially affected ventilators. Covidien is updating software that will modify ventilator alarms and status display messages during a GUI reset. The Puritan Bennett 980 Ventilator Operator Manual will be updated to delete,replace ventilator during a GUI transient reset, and to inform users to monitor the secondary status display until the primary display refreshes. Covidien will communicate with customers when the software and operator manual updates are available. Customers with questions were instructed to contact Technical Support Department at 800-255-6774, option 4, and then option 1. For questions regarding this recall call 303-876-8870.
Quantity in commerce
418 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Product codes and lots
All Puritan Bennett 980 ventilators

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1058-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1058-2015Class IITerminatedVoluntary: Firm initiated