Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1052-2015Class II

Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00...

Zimmer GmbH / initiated 2014-12-11 / Terminated / root cause: design

Zimmer, Inc. began a recall of Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04. Product Usage: The Persona... on . FDA classified it as Class II and gives the reason as "Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1052-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04. Product Usage: The Persona TASPs are sterilizable instruments Intended for multiple uses. The TASP components are used to aid the surgeon in assessing the thickness of the articular surface required. The TASP construct is utilized while conducting a trial range of motion. The TASP consists of 3 parts: a TASP bottom, a TASP shim, and a TASP top. The shims come in varying thicknesses to allow trialing for specific Articular Surface implant thicknesses.
Product code
OIY Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive
Reason for recall
Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22.
Root cause tag
design
FDA root cause
Device Design
Action
Zimmer sent an URGENT MEDICAL DEVICE CORRECTION notifications letter dated December 11, 2014 to affected distributors via electronic mail, and all hospital Risk Managers and distributors with affected inventory via courier. The letter identified the affected product, problem, instructions for responding to the formal recall notification.and actions to be taken. For questions call the customer call center at 1-800-348-2759.
Quantity in commerce
67963
Distribution
Worldwide Distribution - USA Nationwide including DC, and PR and the countries of Australia, Austria, Belgium, Canada, Czech Republic, Dominican Republic, France, Germany, India, Iran, Israel, Italy, Japan, Luxembourg, Netherlands, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Product codes and lots
Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04: 42-5279-002-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 10mm; 42-5279-002-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 12mm; 42-5279-002-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 14mm; 42-5279-004-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 10mm; 42-5279-004-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 12mm; 42-5279-004-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 14mm; 42-5279-006-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 10mm; 42-5279-006-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 12mm; 42-5279-006-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 14mm; 42-5279-008-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 10mm; 42-5279-008-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 12mm; 42-5279-008-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 14mm; 42-5279-003-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 10mm; 42-5279-003-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 11mm; 42-5279-003-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 12mm; 42-5279-003-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 13mm; 42-5279-003-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 14mm; 42-5279-005-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 10mm; 42-5279-005-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 11mm; 42-5279-005-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 12mm; 42-5279-005-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 13mm; 42-5279-005-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 14mm; 42-5279-007-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 10mm; 42-5279-007-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 11mm; 42-5279-007-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 12mm; 42-5279-007-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 13mm; 42-5279-007-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 14mm.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1052-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1052-2015Class IITerminatedVoluntary: Firm initiated