Z-1147-2015Class II

The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the...

Medtronic Neuromodulation / initiated 2015-01-09 / Terminated / root cause: labeling and UDI

Medtronic Neuromodulation began a recall of The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral... on . FDA classified it as Class II and gives the reason as "Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1147-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or muscles when connected to an extension and/or lead. The Medtronic InterStim II Neurostimulator, Model 3058, and Medtronic InterStim Neurostimulator, Model 3023, are parts of neurostimulation systems for InterStim Therapy. The neurostimulators are multiprogrammable devices that accommodates a lead through which a stimulation program is delivered. The Medtronic Enterra Neurostimulator, Model 3116, is part of an Enterra Therapy system for gastric electrical stimulation (GES). The neurostimulator is a multiprogrammable device that accommodates a lead through which a stimulation program is delivered.
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Consignees were sent a Medtronic {Medical Device Correction" letter dated January 9, 2015. The letter was addressed to Healthcare Professional. The letter described the issue and the product involved in the the recall. The letter also listed the Occurrence of the problem and labeling update. A Physician Reply Form was included which requested consignees to return the form to Medtronics. For questions they can contact Medtronic Neuromodulation Technical Services at 1-8007070944 weekdays 7 am 6 pm Central time. A second "Physician Notification Confirmation Form" was included to be completed by Medtronic Representatives.
Quantity in commerce
396,975 (258,805 USA, 138,170 OUS)
Distribution
Worldwide Distributon- USA( nationwide including DC and PR) and Internationally to Mexico and Canada
Product codes and lots
All lot/serial numbers are affected for all 4 Model numbers:  " Model 7425  " Model 3058  " Model 3023  " Model 3116

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1147-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1147-2015Class IITerminatedVoluntary: Firm initiated