Z-1047-2015Class II
Fresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time...
Fresenius Medical Care Renal Therapies Group, LLC / initiated 2014-12-19 / Terminated / root cause: design
Fresenius Medical Care Holdings, Inc. began a recall of Fresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time monitor for non-invasive hematocrit, oxygen... on . FDA classified it as Class II and gives the reason as "Potential for misinterpretation of the graphic display of the Blood Volume ( BV) slope". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1047-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Fresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time monitor for non-invasive hematocrit, oxygen saturation and percent change in blood volume calculation during hemodialysis treatment.
- Product code
- KOC Accessories, Blood Circuit, Hemodialysis
- Reason for recall
- Potential for misinterpretation of the graphic display of the Blood Volume ( BV) slope
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- FMCRTG, LLC representatives contacted Clinics on 12/19/14 by visit/telephone phone) and removed all CLiC units.
- Quantity in commerce
- 35 units
- Distribution
- CT, NY
- Product codes and lots
- Serial Numbers: 1C31M140038 1C31M140040 1C31M140041 1C31M140042 1C31M140033 1C31M140034 1C31M140035 1C31M140030 1C31M140075 1C31M140031 1C31M140027 1C31M140054 1C31M140057 1C31M140060 1C31M140039 1C31M140056 1C32M140005, 1C32M140059, 1C32M140058, 1C32M140014, 1C32M140013, 1C32M140004, 1C32M140012, 1C32M140011, 1C32M140009, 1C32M140008, 1C32M140003, 1C32M140007, 1C32M140006, 1C32M140023, 1C32M140021, 1C32M140022, 1C32M140020, 1C31M140073, 1C31M140050
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1047-2015 | Class II | Terminated | Voluntary: Firm initiated |