Z-1047-2015Class II

Fresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2014-12-19 / Terminated / root cause: design

Fresenius Medical Care Holdings, Inc. began a recall of Fresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time monitor for non-invasive hematocrit, oxygen... on . FDA classified it as Class II and gives the reason as "Potential for misinterpretation of the graphic display of the Blood Volume ( BV) slope". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1047-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time monitor for non-invasive hematocrit, oxygen saturation and percent change in blood volume calculation during hemodialysis treatment.
Product code
KOC Accessories, Blood Circuit, Hemodialysis
Reason for recall
Potential for misinterpretation of the graphic display of the Blood Volume ( BV) slope
Root cause tag
design
FDA root cause
Device Design
Action
FMCRTG, LLC representatives contacted Clinics on 12/19/14 by visit/telephone phone) and removed all CLiC units.
Quantity in commerce
35 units
Distribution
CT, NY
Product codes and lots
Serial Numbers: 1C31M140038 1C31M140040 1C31M140041 1C31M140042 1C31M140033 1C31M140034 1C31M140035 1C31M140030 1C31M140075 1C31M140031 1C31M140027 1C31M140054 1C31M140057 1C31M140060 1C31M140039 1C31M140056 1C32M140005, 1C32M140059, 1C32M140058, 1C32M140014, 1C32M140013, 1C32M140004, 1C32M140012, 1C32M140011, 1C32M140009, 1C32M140008, 1C32M140003, 1C32M140007, 1C32M140006, 1C32M140023, 1C32M140021, 1C32M140022, 1C32M140020, 1C31M140073, 1C31M140050

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1047-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1047-2015Class IITerminatedVoluntary: Firm initiated