The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided...
Brainlab AG began a recall of The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive... on . FDA classified it as Class II and gives the reason as "The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1215-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Brainlab, Inc.
- Product
- The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive orthopedic surgery. (a stereotaxic instrument)
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.
- Root cause tag
- sterility
- FDA root cause
- Reprocessing Controls
- Action
- FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letters dated December 11, 2014 were sent to all direct accounts (customers). The letters included instructions for customers to utilize the sterilization cycles included in the letter. Customers were further advised to alert users of these devices to the sterilization process changes. Customers with questions can contact Brainlab by telephone at 800-597-5911 or via e-mail at [email protected].
- Quantity in commerce
- 114 offset cup impactors
- Distribution
- Worldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom
- Product codes and lots
- Model 52856 BL CURVED CUP INSERTER UNIVERSAL SHAFT; and MODEL 52856A BL OFFSET CUP IMPACTOR UNIVERSAL SHAFT, all serial/lot numbers distributed prior to January 13, 2015.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1215-2015 | Class II | Terminated | Voluntary: Firm initiated |