Z-1215-2015Class II

The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided...

Brainlab, Inc. / initiated 2015-01-16 / Terminated / root cause: sterility

Brainlab AG began a recall of The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive... on . FDA classified it as Class II and gives the reason as "The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1215-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Brainlab, Inc.
Product
The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive orthopedic surgery. (a stereotaxic instrument)
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.
Root cause tag
sterility
FDA root cause
Reprocessing Controls
Action
FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letters dated December 11, 2014 were sent to all direct accounts (customers). The letters included instructions for customers to utilize the sterilization cycles included in the letter. Customers were further advised to alert users of these devices to the sterilization process changes. Customers with questions can contact Brainlab by telephone at 800-597-5911 or via e-mail at [email protected].
Quantity in commerce
114 offset cup impactors
Distribution
Worldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom
Product codes and lots
Model 52856 BL CURVED CUP INSERTER UNIVERSAL SHAFT; and MODEL 52856A BL OFFSET CUP IMPACTOR UNIVERSAL SHAFT, all serial/lot numbers distributed prior to January 13, 2015.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1215-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1215-2015Class IITerminatedVoluntary: Firm initiated