510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1066-2015Class II

System 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker...

Stryker Corp. / initiated 2015-01-07 / Terminated / root cause: sterility

Stryker Instruments Div. of Stryker Corporation began a recall of System 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker Non-sterile Battery. Accessory for powered... on . FDA classified it as Class II and gives the reason as "Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1066-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
System 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker Non-sterile Battery. Accessory for powered orthopedic surgical instrument.
Product code
HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
Reason for recall
Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker.
Root cause tag
sterility
FDA root cause
Under Investigation by firm
Action
Stryker sent notice to customers via Fed ex overnight on 1/16/2015, regarding its voluntary recall of Stryker System 6 Aseptic Housings because the laser welder may have failed to produce a continuous bond line for the four specific lots: 13205, 13209, 13210, 13212. Customers are instructed to discontinue use, complete the business reply form, and return all recalled product. Customers may contact Kara Spath at 269-389-4518, or [email protected].
Quantity in commerce
399
Distribution
Worldwide Distribution-US (nationwide) including Puerto Rico and the states of AR, CA, CO, FL, GA, HI, MD, NC, NY, PA, SC, TX, and VA, and the countries of Australia, Canada, Switzerland, Hong Kong, South Africa, Chile, Korea, Poland, India, Colombia, The Netherlands, and France.
Product codes and lots
Part No. 6126-120-000, Lots 13205, 13209, 13210, 13212

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1066-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1066-2015Class IITerminatedVoluntary: Firm initiated