System 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker...
Stryker Instruments Div. of Stryker Corporation began a recall of System 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker Non-sterile Battery. Accessory for powered... on . FDA classified it as Class II and gives the reason as "Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1066-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- System 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker Non-sterile Battery. Accessory for powered orthopedic surgical instrument.
- Product code
- HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
- Reason for recall
- Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker.
- Root cause tag
- sterility
- FDA root cause
- Under Investigation by firm
- Action
- Stryker sent notice to customers via Fed ex overnight on 1/16/2015, regarding its voluntary recall of Stryker System 6 Aseptic Housings because the laser welder may have failed to produce a continuous bond line for the four specific lots: 13205, 13209, 13210, 13212. Customers are instructed to discontinue use, complete the business reply form, and return all recalled product. Customers may contact Kara Spath at 269-389-4518, or [email protected].
- Quantity in commerce
- 399
- Distribution
- Worldwide Distribution-US (nationwide) including Puerto Rico and the states of AR, CA, CO, FL, GA, HI, MD, NC, NY, PA, SC, TX, and VA, and the countries of Australia, Canada, Switzerland, Hong Kong, South Africa, Chile, Korea, Poland, India, Colombia, The Netherlands, and France.
- Product codes and lots
- Part No. 6126-120-000, Lots 13205, 13209, 13210, 13212
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1066-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY