510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1115-2015Class II

GemLock Long Hex Driver, Catalog RHL2.5; Also distributed within the Tapered SwissPlus & SwissPlus Implant...

ZimVie US Corp LLC / initiated 2015-01-22 / Terminated / root cause: design

Zimmer Dental Inc began a recall of GemLock Long Hex Driver, Catalog RHL2.5; Also distributed within the Tapered SwissPlus & SwissPlus Implant Systems Kits, Surgical, Complete, Catalog OPCST... on . FDA classified it as Class II and gives the reason as "The Gemlock Long Hex Driver may not fit into the Fixture Mount Transfer (FMT) or the internal hex of the implant". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1115-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
GemLock Long Hex Driver, Catalog RHL2.5; Also distributed within the Tapered SwissPlus & SwissPlus Implant Systems Kits, Surgical, Complete, Catalog OPCST; Tapered Screw-Vent Implant System Surgical Kit, Complete, Catalog TSVKIT. Used as a manual delivery tool that engages the Fixture Mount Transfer (FMT) to transport the dental implant from the product packaging to the osteotomy.
Product code
NDP Accessories, Implant, Dental, Endosseous
Reason for recall
The Gemlock Long Hex Driver may not fit into the Fixture Mount Transfer (FMT) or the internal hex of the implant.
Root cause tag
design
FDA root cause
Device Design
Action
An Urgent Device Recall Notice was sent on 1/22/15 to their domestic and foreign customers who purchased the Gemlock Long Hex Driver. The letter informs the customers of the problems identified, risks, and the actions to be taken. Customers are instructed to review the notice and check their inventory for the affected units. All affected product is to be immediately quarantined and removed from service. Customers are to contact Simmer Dental Customer Service at 1-800-854-7019 to obtain a Return Authorization Number and set up return of the products. Customers should complete the attached Business Reply Form and fax it to (574) 372-4265 or e-mail to [email protected]. Customers with questions are instructed to contact Zimmer Dental Customer Service at (800) 854-7019, Mon-Fri, 7 am-5pm (Pacific). Zimmer Dental is expanding the recall to include 1 additional lot for catalog no.TSVKIT. Zimmer Dental plans to inform an amended customer notification letter to their US and foreign consignees. The letter will go out to customers on 4/20/15 to inform them of the expanded lot.
Quantity in commerce
627 units
Distribution
Worldwide Distribution -- US, Argentina, Australia, Bulgaria, Canada, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Egypt, France, Germany, Hong Kong, Hungary, Iran, Israel, Italy, Japan, Lebanon, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Pakistan, Panama, Poland, Romania, Russia, Saudi Arabia, Serbia, Taiwan, Thailand, Tunisia, Turkey, and UAE. Expanded Distribution: Spain
Product codes and lots
Etched Lot No. 62699693 & 62735177. Lot Number on Label: 62712431, 62743168, 62766610, 62702589, 62710484, 62712437, 62743075, 62750487, 62753787, 62756920, 62759862, 62760388, 62768456, and 62773158.  Expanded Lot No. 62728173.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1115-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1115-2015Class IITerminatedVoluntary: Firm initiated