510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1241-2015Class II

LFT v40 Femoral Head; Catalog Number 6260-9-032; V40 CoCr LFit Head 32mm/-4. Designed to be assembled onto a...

Howmedica Osteonics Corp / initiated 2015-02-03 / Terminated

Stryker Howmedica Osteonics Corp. began a recall of LFT v40 Femoral Head; Catalog Number 6260-9-032; V40 CoCr LFit Head 32mm/-4. Designed to be assembled onto a femoral hip stem trunnion during surgery for a... on . FDA classified it as Class II and gives the reason as "Three customer complaints were received for the same batch of V40 LFIT Vitallium femoral heads (part number 6260-9-032, lot 48681201) alleging that the femoral head couldn't be assembled with its corresponding V40 stem...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1241-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
LFT v40 Femoral Head; Catalog Number 6260-9-032; V40 CoCr LFit Head 32mm/-4. Designed to be assembled onto a femoral hip stem trunnion during surgery for a total hip arthroplasty.
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
Three customer complaints were received for the same batch of V40 LFIT Vitallium femoral heads (part number 6260-9-032, lot 48681201) alleging that the femoral head couldn't be assembled with its corresponding V40 stem trunnion.
FDA root cause
Under Investigation by firm
Action
Branches/Agencies were notified of this action by e-mail on December 19, 2014 and were asked to quarantine affected devices. Notification Letters and Product Accountability Forms were sent to Hospital Risk Management and Chief of Orthopaedics via FedEx (with return receipt) by 2/3/2015.
Quantity in commerce
9 units
Distribution
Distributed in the states of GA, MA, NC, NY & PA.
Product codes and lots
Lot Code 48681201

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1241-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1241-2015Class IITerminatedVoluntary: Firm initiated