Hemodialysis Catheters and PICCs: Product identification: MCDLTSL24P (SF X 24CM LT HEMO-CATH), MC061422...
Medical Components, Inc dba MedComp began a recall of Hemodialysis Catheters and PICCs: Product identification: MCDLTSL24P (SF X 24CM LT HEMO-CATH), MC061422 (14FX15CM SLX DBL LMN ST CATH), MRCTP41017 (4F X 55CM... on . FDA classified it as Class II and gives the reason as "Drug products contained within the kits may have been rendered ineffective by a duplicate sterilization process. The kits include one or more of the following drugs: 5ml ampule of lidocaine, 3ml Chloraprep one-step...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1083-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medical Components, Inc.
- Product
- Hemodialysis Catheters and PICCs: Product identification: MCDLTSL24P (SF X 24CM LT HEMO-CATH), MC061422 (14FX15CM SLX DBL LMN ST CATH), MRCTP41017 (4F X 55CM SINGLE CT PICC), MRCTP52007 (5F DUAL PRO-PICC NURSING TRAY), MRCTP52010 (5F DUAL PRO-PICC MAX BARR TRAY), MR81013106MB (3FX60CM SGL TAPRLS VASCU-PICC). Intended uses: To attain long-term vascular access for hemodialysis and apheresis. The Medcomp SLX Silicone Double Lumen Catheter is designed for acute hemodialysis and apheresis. The Pro-PICC CT catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media, and allows for central venous pressure monitoring. The Peripherally Inserted Central Vein Access Catheters are designed for Long or Short- Term peripheral access to the central venous system for intravenous therapy and blood sampling, and allows for central venous pressure monitoring.
- Product code
- LFJ Catheter, Subclavian
- Reason for recall
- Drug products contained within the kits may have been rendered ineffective by a duplicate sterilization process. The kits include one or more of the following drugs: 5ml ampule of lidocaine, 3ml Chloraprep one-step, 10ml pre-filled saline syringe, Hygea saline wipes, and 3ml tinted Chloraprep. No injuries or illnesses were reported.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Medcomp sent a Product Alert/Recall letter dated January 21, 2015 to customers via email. The letter identified the affected product, problem and actions to be taken. Medcomp requested the return of all affected product. Customers were instructed to contact their customer service representative for instructions for returning the product.
- Quantity in commerce
- 243
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Lot numbers: MBWK970, MBWH890, MBWJ180, MBWJ460, MBWH910, MBWH940
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1083-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY