Z-1130-2015Class II

Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological...

Haemonetics Corporation / initiated 2015-01-26 / Terminated / root cause: manufacturing process

Haemonetics Corporation began a recall of Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG)... on . FDA classified it as Class II and gives the reason as "Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1130-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Haemonetics Corporation
Product
Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.
Product code
JPA System, Multipurpose For In Vitro Coagulation Studies
Reason for recall
Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Haemonetics issued on 1/26/15, an Urgent Medical Device Recall to consignees The letter identified the affected product, the reason for the recall, as well as the risk to health. It instructs consignees to inspect their stock and determine if they have any of the affected lots. Customers are to immediately discontinue use and discard as instructed any affected product. The attached customer acknowledgment form should be completed and returned whether or not customers have any affected product. Customers are to contact their local service representative at 1-800-537-2802 if replacement product is needed. If any affected product has been forwarded to other locations, customers are to inform them of the recall. Please call the Technical Support team at 1-800-438-2834, with questions about this action.
Quantity in commerce
4,143 kits
Distribution
Worldwide Distribution -- US, Canada, Czech Republic, Denmark, Great Britain, and Sweden.
Product codes and lots
Lot numbers: HMO 9200  HMO 9217  HMO 9226  HMO 9232  HMO 9276  HMO 9285  HMO 9292 Exp. Date 1/2017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1130-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1130-2015Class IITerminatedVoluntary: Firm initiated