Z-1063-2015Class II

ACUSON S2000 Automated Breast Volume Scanner (ABVS), model number 10434730. The affected sub-components are...

Siemens Medical Solutions USA, Inc. / initiated 2015-01-12 / Terminated / root cause: supplier

Siemens Medical Solutions USA, Inc. began a recall of ACUSON S2000 Automated Breast Volume Scanner (ABVS), model number 10434730. The affected sub-components are #10434968 - Boom Welding and 10434969 - Arm... on . FDA classified it as Class II and gives the reason as "There is a low probability the scanner arm will become completely detached from the scanner column". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1063-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ACUSON S2000 Automated Breast Volume Scanner (ABVS), model number 10434730. The affected sub-components are #10434968 - Boom Welding and 10434969 - Arm Welding. The S2000 ABVS ultrasound System for imaging the breast allows volumetric acquisition of the breast image.
Product code
IYO System, Imaging, Pulsed Echo, Ultrasonic
Reason for recall
There is a low probability the scanner arm will become completely detached from the scanner column.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
Siemens sent a Customer Safety Advisory Notification to all affected sites on January 12, 2015, certified mail, with return receipt. The letter identified the affected product, the issue, and potential risk. Customers were asked to immediately discontinue use of the affected product. A Customer Service Engineer will contact customers to schedule an appointment to make the necessary repairs. Questions should be directed to a local service support person for information regarding timelines and status. Customers are to share the information with all personnel, who need to be aware of the issue.
Quantity in commerce
24 units
Distribution
Worldwide Distribution -- US, including the states of TX and MN, and the territory of P.R.; and, the countries of China, Germany, Russian Fed, Saudi Arabia, Taiwan, France, Thailand, and Italy.
Product codes and lots
Serial numbers 250157-250190.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1063-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1063-2015Class IITerminatedVoluntary: Firm initiated