Z-0346-2016Class II
Philips Healthcare Allura Xper Series: Allura Xper FD10 Allura Xper FD10/10 Allura Xper FD20 Allura Xper FD21...
Philips North America LLC / initiated 2015-02-03 / Terminated / root cause: manufacturing process
Philips Medical Systems, Inc. began a recall of Philips Healthcare Allura Xper Series: Allura Xper FD10 Allura Xper FD10/10 Allura Xper FD20 Allura Xper FD21 Allura Xper FD22 Allura Xper FD23 Allura Xper... on . FDA classified it as Class II and gives the reason as "Monitor Ceiling Suspension system may fall". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0346-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Healthcare Allura Xper Series: Allura Xper FD10 Allura Xper FD10/10 Allura Xper FD20 Allura Xper FD21 Allura Xper FD22 Allura Xper FD23 Allura Xper FD24 Allura Xper FD25 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Monitor Ceiling Suspension system may fall
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Philips sent an Urgent Field Safety Notice letter dated January 8, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. The actuator of all affected systems will be replaced free of charge. All affected customers will be contacted in order to schedule the implementation of this action. For questions contact your local Philips representatice: Technical Support Line: 1-800-722-9377.
- Quantity in commerce
- 9 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
- Product codes and lots
- System : S/N : Model: 722003 30 Allura Xper FD10 722010 670 Allura Xper FD10 722011 167 Allura Xper FD10/10 722012 1951 Allura Xper FD20 722012 536 Allura Xper FD21 722012 504 Allura Xper FD22 722012 748 Allura Xper FD23 722012 536 Allura Xper FD24 722012 579 Allura Xper FD25
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0346-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0347-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0348-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0349-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0350-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0351-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0352-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0353-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0354-2016 | Class II | Terminated | Voluntary: Firm initiated |