Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1281-2015Class II

Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number...

Synthes GmbH / initiated 2015-02-17 / Terminated / root cause: manufacturing process

Synthes, Inc. began a recall of Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for... on . FDA classified it as Class II and gives the reason as "For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1281-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.
Product code
FZX Guide, Surgical, Instrument
Reason for recall
For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An Urgent Notice: Medical Device Recall, dated February 17, 2015, was sent to all direct consignees and sales consultants instructing them to take the following actions: Take note of the quantity of affected parts in your possession, and then follow ONE of the options for replacement of the affected product: Option 1: Call DePuy Synthes Repair Department at 1-800-288-6698 to schedule a repair for the affected products. Complete the Verification Section following the instructions provided and send a copy of the completed Verification Section to DePuy Synthes by: Fax: 866-723-2823 or Scan/e-mail: [email protected] Option 2: Call DePuy Synthes Repair Department at 1-800-288-6698 to order repair part. The replacement part will be sent directly to your facility. Questions should be directed to 610-719-5450 or a local DePuy Synthes sales consultant.
Quantity in commerce
183
Distribution
Nationwide Distribution
Product codes and lots
Part numbers:03.010.048 and 03.010.052 with  Lot Numbers:  4874015 4874021 4927262 4954481 4963192 4874016 4874022 4944750 4957586 4963193 4784017 4899564 4944751 4957587 4972003 4874018 4899565 4954478 4957588 4972004 4874019 4927260 4954479 4957589  4874020 4927261 4954480 4963190  4922533 4922975 4981158  4922970 4923015 4981159  4922971 4977241 4981160  4922972 4977242 4981162  4922973 4981156 5035905  4922974 4981157 5035906

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1281-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1281-2015Class IITerminatedVoluntary: Firm initiated