Zip 8i Surgical Skin Closure Device; for incisions up to 8 cm. Product number PS2080 Indicated for use during...
Zipline Medical began a recall of Zip 8i Surgical Skin Closure Device; for incisions up to 8 cm. Product number PS2080 Indicated for use during and after skin incision procedures to approximate... on . FDA classified it as Class II and gives the reason as "The manufacturers seal may not be present on the Nylon pouch resulting in lack of assurance for sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1234-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zipline Medical, Inc.
- Product
- Zip 8i Surgical Skin Closure Device; for incisions up to 8 cm. Product number PS2080 Indicated for use during and after skin incision procedures to approximate skin and hold together skin edges until healing can take place.
- Product code
- KGX Tape And Bandage, Adhesive
- Reason for recall
- The manufacturers seal may not be present on the Nylon pouch resulting in lack of assurance for sterility.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Zipline sent an Urgent Voluntary product Recall letters to customers on February 5, 2015 by e-mail with possible follow up on site by firm representatives. The letter identified the affected product, problem and actions to be taken. Customers are asked to quarantine inventory immediately and examine the affected lots for complete pouch seals. Any product found with an incomplete seal should be destroyed or returned to ZipLine Medical. Please complete the enclosed response form as soon as possible, even if you no longer have stock of the lot numbers subject to this correction. This information is required to reconcile the inventory. Please contact Customer Service at 1-888-326-8999 to coordinate any product returns, replacement stock, or credit notes.
- Quantity in commerce
- 3109 - total - all model numbers
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Panama, Tehran, Ireland, Sweden, Japan, Hong Kong, Finland, Pakistan, Taiwan, Malaysia.
- Product codes and lots
- Lot 1002690, Exp 2015-08
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1234-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1235-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1236-2015 | Class II | Terminated | Voluntary: Firm initiated |