ADVIA Centaur Systems Progesterone Kit, (1 -pack) In Vitro Diagnostic; Catalog number: 02382928; SMN: 10310305
Siemens Healthcare Diagnostics, Inc began a recall of ADVIA Centaur Systems Progesterone Kit, (1 -pack) In Vitro Diagnostic; Catalog number: 02382928; SMN: 10310305 on . FDA classified it as Class II and gives the reason as "ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL (95.4 nmol/L)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1229-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- ADVIA Centaur Systems Progesterone Kit, (1 -pack) In Vitro Diagnostic; Catalog number: 02382928; SMN: 10310305.
- Product code
- JLS Radioimmunoassay, Progesterone
- Reason for recall
- ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL (95.4 nmol/L)
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Siemens issued an Urgent Medical Device Correction (UMDC) was sent to affected Siemens Healthcare Diagnostics customers in the United States on January 27, 2015 to be delivered to customers on January 28, 2015. An Urgent Field Safety Notice (U FSN) was sent out for distribution to all customers outside the United States on January 27, 2015. These notices inform the customer of the potential to observe a high bias on patient sample with results greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. To mitigate the bias, Siemens instructs customers to dilute samples with values greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. In addition, Siemens informs customers to continue to follow Urgent Medical Device Correction 10819675, Rev. A and Urgent Field Safety notice 10819674, Rev A (Information Regarding the ADVIA Centaur Systems Calibrator E), dated September 2014. Fax-back forms to be returned to confirm that the customers received the Urgent Medical Device Correction.
- Quantity in commerce
- 9778 kits
- Distribution
- Worldwide Distribution-US (nationwide), Austria, Bahrain, Belgium, Bulgaria, Brazil, Canada, Chile, Colombia, Croatia, Cyprus, Czech, Republic, Denmark, Egypt, Estonia, France, Georgia, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Serbia, Slovakia, Spain, South America, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and Vatican.
- Product codes and lots
- Lot Numbers: 42564268 42621268 43834268 Exp. Date: May 22, 2015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1229-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1230-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1231-2015 | Class II | Terminated | Voluntary: Firm initiated |