Apollo (cold) Laser Desktop Control Units, Model AP2-DT. The Apollo IR Heat Lamp System is intended to emit...
Pivotal Health Solutions began a recall of Apollo (cold) Laser Desktop Control Units, Model AP2-DT. The Apollo IR Heat Lamp System is intended to emit energy in the infrared spectrum to provide topical... on . FDA classified it as Class II and gives the reason as "Control units were equipped with an internal mounting kit that does not meet medical safety standards, and are conductive, increasing the risk of electric shock to the user and patient. These units were manufactured...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1250-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Pivotal Health Solutions
- Product
- Apollo (cold) Laser Desktop Control Units, Model AP2-DT. The Apollo IR Heat Lamp System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, the temporary increase in local blood circulation and / or promoting relaxation of muscle.
- Product code
- ILY Lamp, Infrared, Therapeutic Heating
- Reason for recall
- Control units were equipped with an internal mounting kit that does not meet medical safety standards, and are conductive, increasing the risk of electric shock to the user and patient. These units were manufactured prior to Pivotal Health Solutions acquisition of the Apollo product line.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, Pivotal Health Solutions, sent a letter dated December 17, 2014 on 12/17/2014 and an amended Pivotal "Urgent Medical Device Recall" letter dated February 9, 2015 to its consignees/customers. The letters described the product, problem and actions to be taken. The consignees/customers were instructed to stop using the unit and contact Pivotal Health Solution's at 1-800-743-7738 to arrange for return and repair asap; immediately examine your device inventory and quarantine any product subject to recall; if you have further distributed the product, identify your customers and notify them at once of this product recall, and complete and return the enclosed RETURN AUTHORIZATION FORM with the units to Pivotal Service Center, 1654 Mardon Drive, Dayton, OH 45432 and the DECLARATION OF CONTAMINATION STATUS form via Fax to: 605-882-8398. If you have any questions, contact Pivotal Health Solutions Service Repair Coordinator at 800-743-7738 or email [email protected].
- Quantity in commerce
- 35
- Distribution
- US Distribution to states of: AZ, AR, CA, GA, IL, IN, IA, KS, MI, MN, MO, NY, OH, OR, PA, UT and WA.
- Product codes and lots
- DT-1102, 1105, 1106, 1109, 1110, 1112,1114, 1115, 1116, 1118,1119, 1120, 1121, 1121, 1122, 1123, 1124, 1125, 1126, 1127, 1128, 1129, 1130, 1131, 1132, 1134, 1135, 1136, 1137, 1138, 1139, 1302, 1303, 1304, 1309.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1250-2015 | Class II | Terminated | Voluntary: Firm initiated |