TactiSys Quartz Pack, PN-004 400. For percutaneous catheter radiofrequency (RF) ablation of atrial cardiac...
St Jude Medical began a recall of TactiSys Quartz Pack, PN-004 400. For percutaneous catheter radiofrequency (RF) ablation of atrial cardiac arrhythmias that allows visualization of the contact... on . FDA classified it as Class II and gives the reason as "A configuration update needs to be done on TactiSys to appropriately recognize all TactiCath catheters". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1284-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- TactiSys Quartz Pack, PN-004 400. For percutaneous catheter radiofrequency (RF) ablation of atrial cardiac arrhythmias that allows visualization of the contact force between TactiCath quartz catheter tip and the heart wall. Product Usage: This device is indicated for use in cardiac electrophysiological mapping and for the treatment of drug refractory recurrent symptomatic paroxysmal atrial fibrillation, when used in conjunction with a compatible radiofrequency generator and three-dimensional mapping system.
- Product code
- OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation
- Reason for recall
- A configuration update needs to be done on TactiSys to appropriately recognize all TactiCath catheters.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- St. Jude Medical sent a Service Bulletin to inform of a TactiSys Configuration File update that needs to be performed. A technical service personnel will visit each consignee and performed the software change on-site. For questions about this Service Bulletin, contact Technical Support ([email protected]).
- Quantity in commerce
- 89
- Distribution
- US Nationwide Distribution in the states of: AL, AZ, CA, CO, CT, FL, GA, ID, IL IN, KS, LA, MD, MA, MI, MN,MS, NE, NV,NH, NM, NY, NC, OH, OK OR, PA, TN, TX, UT, VA,WA.
- Product codes and lots
- 30078, 30118, 30368, 30059, 30098, 30169, 30110, 30042, 30108, 30066, 30067, 30117, 30172, 30063, 30061, 30060, 30161, 30183, 30129, 30097, 30341, 30043, 30344, 30139, 30112, 30148, 30071, 30096, 30221, 30178, 30204, 30076, 30332, 30343, 30056, 30082, 30123, 30091, 30054, 30087, 30218, 30173, 30143, 30095, 30157, 30176, 30068, 30152, 30158, 30080, 30083, 30153, 30093, 30452, 30121, 30160, 30165, 30088, 30180, 30195, 40000, 30109, 30090, 30147, 30156, 30174, 30124, 30058, 30105, 30070, 30170, 30084, 30333, 30089, 30350, 30222, 30228, 30229, 30134, 30045, 30346, 30177, 30184, 30454, 30111, 30075, 30073, 30219, 30085.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1284-2015 | Class II | Terminated | Voluntary: Firm initiated |