Z-1116-2015Class II

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that...

Terumo Cardiovascular Systems Corporation / initiated 2015-02-06 / Terminated / root cause: design

Terumo Cardiovascular Systems Corporation began a recall of Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood... on . FDA classified it as Class II and gives the reason as "Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1116-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.
Product code
DWC Controller, Temperature, Cardiopulmonary Bypass
Reason for recall
Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.
Root cause tag
design
FDA root cause
Device Design
Action
On 2/6/2015, URGENT MEDICAL DEVICE RECALL notifications were mailed to the affected hospitals. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call customer service at 1-800-521-2818.
Quantity in commerce
103 units
Distribution
US (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.
Product codes and lots
HX2 Temperature Management System, 120V  Catalog # 809810  serial numbers: 00016, 0001 1116  HX2 Temperature Management System, 100V Catalog # 812231  serial numbers: 0001, 1002, PT0003, PT0004   HX2 Temperature Management System, 240V Catalog # 812240  serial numbers: 0001, 0002  HX2 Temperature Management System, Trial Kit Catalog # 809810TK  serial numbers: 1001 1042

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1116-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1116-2015Class IITerminatedVoluntary: Firm initiated