510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1219-2015Class II

Zimmer Segmental System (ZSS) Cemented Stem / ZSS Cemented Stem, Smooth Stem Prosthesis, Hip...

Zimmer GmbH / initiated 2015-02-05 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of Zimmer Segmental System (ZSS) Cemented Stem / ZSS Cemented Stem, Smooth Stem Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous... on . FDA classified it as Class II and gives the reason as "PMMA coating does not meet specifications. Area of the stem may be missing PMMA coating". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1219-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Segmental System (ZSS) Cemented Stem / ZSS Cemented Stem, Smooth Stem Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous, Uncemented.
Product code
LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Reason for recall
PMMA coating does not meet specifications. Area of the stem may be missing PMMA coating.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Zimmer sent an URGENT MEDICAL DEVICE RECALL - Lot Specific letter dated February 5, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Your Responsibilities 1. Review the notification and ensure affected personnel are aware of the contents. 2. Locate and remove from inventory the affected products identified above. 3. Return any affected product with missing PMMA coating per the PER process 4. Return the Notification Acknowledgment Form (Attachment 1) to [email protected]. 5. Please notify Zimmer if the hospital that you have distributed the affected product to has implanted the product. In addition, identify the surgeons that have used this product. 6. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST.
Quantity in commerce
1 unit
Distribution
US Distribution to the state of TX
Product codes and lots
Item No. 00-5852-052-10; Lot 62866438

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1219-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1219-2015Class IITerminatedVoluntary: Firm initiated