510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1135-2105Class II

CoCr Femoral Head - indicated for hip arthroplasty for noncemented use in skeletally mature invididuals...

Zimmer GmbH / initiated 2014-12-17 / Terminated / root cause: packaging

Zimmer Gmbh began a recall of CoCr Femoral Head - indicated for hip arthroplasty for noncemented use in skeletally mature invididuals undergoing primary surgery for rehabilitating hips... on . FDA classified it as Class II and gives the reason as "Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1135-2105
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
CoCr Femoral Head - indicated for hip arthroplasty for noncemented use in skeletally mature invididuals undergoing primary surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
On 12/17/2014, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.
Quantity in commerce
6109
Distribution
Domestic: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MS, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY International: Canada, Japan
Product codes and lots
Item numbers: 7210-22-000; 7210-22-350; 7210-22-800; 7210-26-000; 7210-26-035; 7210-26-350; 7210-26-800; 7210-28-000; 7210-28-004; 7210-28-400; 7210-28-800; 7210-32-000; 7210-32-004; 7210-32-400; 7210-32-800.  Lot #'s:  61487412, 61516918, 61525215, 61533617, 61582735, 61417397, 61487413, 61417398, 61393996, 61393997, 61417401, 61417402, 61487414, 61487415, 61487416, 61487417, 61487418, 61487420, 61487422, 61487423, 61495211, 61495212, 61500247, 61516884, 61516885, 61516886, 61516888, 61525212, 61525213, 61533606, 61533607, 61533608, 61538022, 61538024, 61538026, 61554502, 61565073, 61565079, 61565081, 61565082, 61582736, 61417403, 61487424, 61487425, 61487426, 61495218, 61516903, 61516904, 61525214, 61533612, 61565084, 61565085, 61577701, 61582737, 61417399, 61487427, 61487428, 61487429, 61495210, 61516881, 61516882, 61533605, 61538021, 61543660, 61565071, 61565072, 61487443, 61565088, 61487447, 61487450, 61500252, 61525210, 61533616, 61565091, 61565104, 61582738, 61487453, 61500245, 61525209, 61565102, 61394048, 61487455, 61500250, 61538042, 61577704, 61487456, 61565101, 61382197, 61394027, 61394029, 61417404, 61417405, 61487571, 61487572, 61487573, 61487574, 61487575, 61487576, 61487577, 61495217, 61516893, 61516894, 61516896, 61516897, 61516898, 61525224, 61525225, 61525226, 61533610, 61538041, 61543661, 61565106, 61565112, 61565119, 61565125, 61382207, 61393990, 61417407, 61487578, 61487579, 61487580, 61487581, 61495237, 61516921, 61516922, 61525230, 61538057, 61543828, 61565116, 61565124, 61582745, 61394050, 61417406, 61487582, 61487583, 61487584, 61500251, 61516927, 61525227, 61525228, 61565109, 61565121, 61402292, 61487585, 61516917, 61525229, 61565122

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1135-2105
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1132-2015Class IITerminatedVoluntary: Firm initiated
Z-1133-2015Class IITerminatedVoluntary: Firm initiated
Z-1134-2015Class IITerminatedVoluntary: Firm initiated
Z-1135-2105Class IITerminatedVoluntary: Firm initiated
Z-1136-2015Class IITerminatedVoluntary: Firm initiated