Metasul Femoral Head - indicated for total hip arthroplasty for treatment of patient conditions of...
Zimmer Gmbh began a recall of Metasul Femoral Head - indicated for total hip arthroplasty for treatment of patient conditions of non-inflammatory degenerative joint disease on . FDA classified it as Class II and gives the reason as "Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1136-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Metasul Femoral Head - indicated for total hip arthroplasty for treatment of patient conditions of non-inflammatory degenerative joint disease.
- Product code
- KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
- Reason for recall
- Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- On 12/17/2014, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.
- Quantity in commerce
- 16
- Distribution
- Domestic: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MS, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY International: Canada, Japan
- Product codes and lots
- Item numbers: 734028800 Lot #'s: 61487614
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1132-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1133-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1134-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1135-2105 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1136-2015 | Class II | Terminated | Voluntary: Firm initiated |