Z-1209-2015Class I

IntellaTip MiFi Open-Irrigated radiofrequency Ablation Catheters: Material number: M004EPM96200, Catalog...

Boston Scientific Corporation / initiated 2015-02-03 / Terminated / root cause: design

Boston Scientific Corp began a recall of IntellaTip MiFi Open-Irrigated radiofrequency Ablation Catheters: Material number: M004EPM96200, Catalog Number: EPM9620: Material number: M004EPM9620K20... on . FDA classified it as Class I and gives the reason as "Reports of formation of char adherent to the proximal part of the distal tip electrodes". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1209-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
IntellaTip MiFi Open-Irrigated radiofrequency Ablation Catheters: Material number: M004EPM96200, Catalog Number: EPM9620: Material number: M004EPM9620K20, Catalog Number: EPM9620K2; Material number : M004EPM9620N40; Catalog number: EPM9620N4. The IntellaTip MiFi Open-Irrigated Ablation Catheter is a 7.5F (2.5 mm) quadrapolar open-irrigated (OI) ablation catheter designed to deliver radiofrequency (RF) energy to the 4.5 mm catheter tip electrode for cardiac ablation The product is not approved for use/sale in the US.
Product code
OAD Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter
Reason for recall
Reports of formation of char adherent to the proximal part of the distal tip electrodes
Root cause tag
design
FDA root cause
Device Design
Action
Customer notification letters were sent by overnight mail on February 10, 2015.
Quantity in commerce
14 units
Distribution
Distributed in Germany, Great Britain, and the Netherlands.
Product codes and lots
Material number: M004EPM96200, Catalog Number EPM9620; Serial numbers: 16941576, 16943730, 17150129, 17156568, 17156570, 17168752, 17171230, 17218387, 17221754, 17237747, 17247469, 17261879.  Material number: M004EPM9620K20, Catalog Number: EPM9620K2: Serial numbers : 17018928, 17018929, 17147617, 17172602, 17175351, 17223957, 17230227, 17233882, 17235011, 17241926, 17263604, 17285884, 17288714.  Material number : M004EPM9620N40; Catalog number: EPM9620N4; Serial numbers: 17148152, 17226126, 17266642.  All with Expiry Dates of April 20, 2015 - June 14, 2015.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1209-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1209-2015Class ITerminatedVoluntary: Firm initiated