510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1140-2015Class II

Maquet CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP)

Datascope Corp. / initiated 2015-01-30 / Terminated

Maquet Datascope Corp - Cardiac Assist Division began a recall of Maquet CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP) on . FDA classified it as Class II and gives the reason as "Power supply malfunction complaints related to suboptimal thermal management". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1140-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Maquet CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP)
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
Power supply malfunction complaints related to suboptimal thermal management.
FDA root cause
Under Investigation by firm
Action
Maquet Inc. sent a recall letter/return response form dated 1/30/2015.
Quantity in commerce
1,300 units
Distribution
US Nationwide distribution
Product codes and lots
0998-00-0800-31 0998-UC-0800-31 0998-00-0800-32 0998-UC-0800-33 0998-00-0800-33 0998-UC-0800-52 0998-00-0800-34 0998-UC-0800-53 0998-00-0800-35 0998-UC-0800-55 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55.   CARDIOSAVE Hybrid IABPs serviced with a new/replacement power supply after August 19, 2014 are not affected.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1140-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1140-2015Class IITerminatedVoluntary: Firm initiated