Z-1280-2015Class II

Passport V Monitor, Multiparameter Patient Monitor(with Arrhythmia Detection and Alarms) Intended for intra...

Datascope Corp. / initiated 2015-01-23 / Terminated / root cause: manufacturing process

Mindray DS USA, Inc. dba Mindray North America began a recall of Passport V Monitor, Multiparameter Patient Monitor(with Arrhythmia Detection and Alarms) Intended for intra hospital use under the direct supervision of a... on . FDA classified it as Class II and gives the reason as "An issue has been identified with Passport V Monitors invasive blood pressure function (IBP) which may provide an incorrect IBP measurement". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1280-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Passport V Monitor, Multiparameter Patient Monitor(with Arrhythmia Detection and Alarms) Intended for intra hospital use under the direct supervision of a licensed healthcare practitioner.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
An issue has been identified with Passport V Monitors invasive blood pressure function (IBP) which may provide an incorrect IBP measurement
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Mindray DS USA Inc sent Urgent: Passport V Monitor Field Corrective Action letters dated 1/23/2015 to hospital administrators. The letters informed customers that the IBP module may provide an incorrect IBP measurement. The letter also explained the adverse effects on patients. Customers are asked to either calibrate the IBP following the instructions provided; or, contact a Mindray Service Representative, who can perform the calibration, at 1-800-288-2121. In the case of the second option, it is recommended that the monitor be removed from service until the calibration is performed. If the first option is chosen, customers are to notify a Service Representative upon completion of the action. Customers can direct their questions to Diane Arpino at 201-995-8407.
Quantity in commerce
519 units
Distribution
Nationwide Distribution
Product codes and lots
6100F- PA00195 6100F- PA00197 6100F- PA00219 6100F- PA00221 6100F- PA00267 6100F- PA00269 6100F- PA00291 6100F- PA00293

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1280-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1280-2015Class IITerminatedVoluntary: Firm initiated