BD Q-Syte Luer Access Split Septum, 0.16 ml, Sterile, Rx only, Catalog REF #385100, Manufacturer: Becton...

Becton Dickinson & Co / initiated 2015-02-19 / Terminated / root cause: manufacturing process

Becton Dickinson & Company began a recall of BD Q-Syte Luer Access Split Septum, 0.16 ml, Sterile, Rx only, Catalog REF #385100, Manufacturer: Becton Dickinson Infusion Therapy Systems Inc. An accessory... on . FDA classified it as Class II and gives the reason as "Becton Dickinson received an elevated number of complaints for blood or infusate/medication leakage due to tearing of the silicone septum". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1317-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Q-Syte Luer Access Split Septum, 0.16 ml, Sterile, Rx only, Catalog REF #385100, Manufacturer: Becton Dickinson Infusion Therapy Systems Inc. An accessory to an intravascular administration set that permits injection, gravity flow or withdrawal of fluids.
Product code
FPA Set, Administration, Intravascular
Reason for recall
Becton Dickinson received an elevated number of complaints for blood or infusate/medication leakage due to tearing of the silicone septum.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Becton Dickinson sent an Urgent Voluntary Product Recall Letter/Recall Response Card/Packing Instructions, dated 2/19/15, to their affected customers via UPS next day mail. The letters identified the affected device and explained the reason for the recall. Customers were to review their inventory for the specific catalog and lot numbers provided and quarantine the affected product. The Recall Response Card form should be completed and faxed back to BD at 1-866-514-2113 or e-mail [email protected]. All affected product should be returned with the completed Recall Response Card following the steps on the enclosed packing instruction. Upon receipt of the returned product, BD will send customers replacement product. If customers have questions or require assistance, they should contact 1-866-800-2917.
Quantity in commerce
315,800 units
Distribution
Distributed to the states of CA, CT, FL, IL, MA, ND, NY, OH and PA.
Product codes and lots
Lot #'s 4128925 and 4128926

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1317-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1317-2015Class IITerminatedVoluntary: Firm initiated