BD Q-Syte Luer Access Split Septum, 0.16 ml, Sterile, Rx only, Catalog REF #385100, Manufacturer: Becton...
Becton Dickinson & Company began a recall of BD Q-Syte Luer Access Split Septum, 0.16 ml, Sterile, Rx only, Catalog REF #385100, Manufacturer: Becton Dickinson Infusion Therapy Systems Inc. An accessory... on . FDA classified it as Class II and gives the reason as "Becton Dickinson received an elevated number of complaints for blood or infusate/medication leakage due to tearing of the silicone septum". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1317-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Q-Syte Luer Access Split Septum, 0.16 ml, Sterile, Rx only, Catalog REF #385100, Manufacturer: Becton Dickinson Infusion Therapy Systems Inc. An accessory to an intravascular administration set that permits injection, gravity flow or withdrawal of fluids.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Becton Dickinson received an elevated number of complaints for blood or infusate/medication leakage due to tearing of the silicone septum.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Becton Dickinson sent an Urgent Voluntary Product Recall Letter/Recall Response Card/Packing Instructions, dated 2/19/15, to their affected customers via UPS next day mail. The letters identified the affected device and explained the reason for the recall. Customers were to review their inventory for the specific catalog and lot numbers provided and quarantine the affected product. The Recall Response Card form should be completed and faxed back to BD at 1-866-514-2113 or e-mail [email protected]. All affected product should be returned with the completed Recall Response Card following the steps on the enclosed packing instruction. Upon receipt of the returned product, BD will send customers replacement product. If customers have questions or require assistance, they should contact 1-866-800-2917.
- Quantity in commerce
- 315,800 units
- Distribution
- Distributed to the states of CA, CT, FL, IL, MA, ND, NY, OH and PA.
- Product codes and lots
- Lot #'s 4128925 and 4128926
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1317-2015 | Class II | Terminated | Voluntary: Firm initiated |