Z-1354-2015Class II

Syngo.plaza. A Picture Archiving and Communication System (PACS) intended to display, process, read, report...

Siemens Medical Solutions USA, Inc. / initiated 2015-02-18 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo.plaza. A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital... on . FDA classified it as Class II and gives the reason as "Possibly incomplete archived studies during pre-fetch. In a server farm setup, when pre-fetch/retrieve operation is performed for partially archived studies, the series that have not yet been archived, will remain...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1354-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo.plaza. A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Possibly incomplete archived studies during pre-fetch. In a server farm setup, when pre-fetch/retrieve operation is performed for partially archived studies, the series that have not yet been archived, will remain unarchived.
Root cause tag
software
FDA root cause
Software design
Action
Safety Advisory Notice, dated February 18, 2015, was sent to end users that identified the product, problem, and action to be taken. Users were advised not to manually remove the delete protection in the STS.
Quantity in commerce
4
Distribution
US Distribution: MA, WI, FL
Product codes and lots
Model Numbers of device: 10592457, 10863172, serial numbers: 100177 100033 100181 100029

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1354-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1354-2015Class IITerminatedVoluntary: Firm initiated