510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1259-2015Class II

Arrow Intra-Aortic Balloon Catheter Kit Fiberoptix (IAB), 8 Fr 40 cc Model: IAB-05840-LWS

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2013-11-08 / Terminated / root cause: manufacturing process

Arrow International Inc began a recall of Arrow Intra-Aortic Balloon Catheter Kit Fiberoptix (IAB), 8 Fr 40 cc Model: IAB-05840-LWS on . FDA classified it as Class II and gives the reason as "Potential for a leak between the catheter and the bifurcate which would allow helium to escape to the atmosphere". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1259-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow Intra-Aortic Balloon Catheter Kit Fiberoptix (IAB), 8 Fr 40 cc Model: IAB-05840-LWS
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
Potential for a leak between the catheter and the bifurcate which would allow helium to escape to the atmosphere.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Arrow International Inc notified accounts by letter dated 11/8/13 describing the problem and requested users to immediately discontinue use and return product. A Recall Acknowledgement form was to be completed and returned.
Quantity in commerce
166 units
Distribution
Distribution US nationwide (including AZ, NJ, ME, TX, WA), and Germany.
Product codes and lots
Lot Number: KF1111796 Exp. Date: 11/30/2013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1259-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1259-2015Class IITerminatedVoluntary: Firm initiated