Z-1398-2015Class II

QuikLYTE Dilution Check: used to monitor and adjust for the dilution ratio on Dimension clinical chemistry...

Siemens Healthcare Diagnostics Inc. / initiated 2015-02-24 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of QuikLYTE Dilution Check: used to monitor and adjust for the dilution ratio on Dimension clinical chemistry systems with the QuikLYTE module. Intended Use of... on . FDA classified it as Class II and gives the reason as "A positive bias beyond 5% resulting in a dilution check failure or an unnecessary dilution check correction if the positive bias is greater than 1% and less than 5%. The dilution check correction would be detected by...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1398-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
QuikLYTE Dilution Check: used to monitor and adjust for the dilution ratio on Dimension clinical chemistry systems with the QuikLYTE module. Intended Use of the QuikLYTE module : The Na+, K+, Cl and TCO2 methods on the Dimension clinical chemistry system with the QuikLYTE module are in vitro diagnostic tests intended for the quantitative measurement of sodium, potassium, chloride and total carbon dioxide in human serum and heparinized plasma. They are also intended for the quantitative measurement of Na+, K+ and Cl- in urine.
Product code
CEM Electrode, Ion Specific, Potassium
Reason for recall
A positive bias beyond 5% resulting in a dilution check failure or an unnecessary dilution check correction if the positive bias is greater than 1% and less than 5%. The dilution check correction would be detected by QC; however, if not detected by QC, the calibration curve would be decreased and cause elevated sodium, potassium and chloride patient results.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
An Urgent Medical Device Recall letter, dated February 2015, was provided to all Dimension system customers who received the affected lot to notify them that Siemens confirmed customer complaints of failing dilution checks when using QuikLYTE Dilution Check lot 4MD707. The letter also instructed them to immediately discontinue use and discard all inventory of Dimension QuikLYTE Dilution Check lot 4MD707. Siemens offered a no charge replacement with a non-impacted lot.
Quantity in commerce
5981
Distribution
Worldwide Distribution.
Product codes and lots
Catalog Number S640, Siemens Material Number (SMN) 10444872, Lot Number 4MD707

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1398-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1398-2015Class IITerminatedVoluntary: Firm initiated