Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1362-2015Class II

GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18...

Orthopediatrics Canada ULC / initiated 2015-02-20 / Terminated

Pega Medical Inc. began a recall of GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18; GAP-N48-26; GAP-N48-28. Product Usage: The GAP... on . FDA classified it as Class II and gives the reason as "Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cortical screw holes. Failures mainly occurred in patients with tibial pseudoarthrosis (CPT) or Osteogenesis...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1362-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthopediatrics Canada ULC
Product
GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18; GAP-N48-26; GAP-N48-28. Product Usage: The GAP Endo-Exo Medullary System is used for the treatment of fractures or correction of deformities in the femur, tibia and humerus of pediatric patients, who have reached skeletal maturity. The GAP Nail is an intramedullary nail. It is the main component of the GAP Endo-Exo Medullary System. The rest of the components of the system are screws, plates and caps that are assembled to the nail during the surgical procedure. The GAP Endo-Exo Medullary System is indicated as a temporary implant to assure alignment, stabilization and fixation of: long bones that have been surgically prepared (osteotomy) for correction of deformities or fractures caused by trauma or disease. The GAP System is used for pediatric patients (child and adolescent) ages 2 to 21. It can be used to correct the following conditions: -Diaphyseal fracture of the femur, tibia and humerus -Fractures of the femoral neck -Subtrochanteric, intertrochanteric and combination fractures -Correction of deformities (OI, Coxa vara, Coxa valga) -Nonunions and malunions
Product code
HSB Rod, Fixation, Intramedullary And Accessories
Reason for recall
Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cortical screw holes. Failures mainly occurred in patients with tibial pseudoarthrosis (CPT) or Osteogenesis imperfecta (OI). Higher failure rates have been detected in tibias with distal fractures that are close to the distal locking holes.
FDA root cause
Other
Action
Pega Medical sent the Urgent Field Safety Notice-Recall/Advisory Notice letter, dated 2015/02/20, to US consignees (surgeons) via email. The following actions were advised: 1. It is recommended to monitor patients as considered necessary in order to ensure that weight bearing does not take place until full consolidation of the fracture/osteotomy is completed and that weight bearing is applied progressively. 2. Report to the distributor or Pega Medical any adverse event observed. In cases where the consolidation has occurred, normal patient follow up can be continued with no serious consequences to the patient. In cases where consolidation has not been achieved after six weeks, replacement of the nail might be necessary. 3. All surgeons receiving this notice must acknowledge reception by a written confirmation sent to the distributor or to the contact reference person indicated below. US consignees with questions can call Pega Medical at 450-688-5144 ext. 242, or send fax to 450-688-1977, or send email to [email protected].
Quantity in commerce
7 units distributed in the US;
Distribution
US Nationwide Distribution in the states of: California, Florida, New York, and Puerto Rico (US).
Product codes and lots
Lot numbers: 110913-001; 110913-002; 110913-004; 110913-005;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1362-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1362-2015Class IITerminatedVoluntary: Firm initiated