Z-1336-2015Class II

Boston Scientific: Expel" Drainage Catheter with Twist-Loc" Hub APDL Drainage Catheter and APDL Drainage...

Boston Scientific Corporation / initiated 2015-02-25 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Boston Scientific: Expel" Drainage Catheter with Twist-Loc" Hub APDL Drainage Catheter and APDL Drainage Catheter Kit, Expel" Drainage Catheter APD Drainage... on . FDA classified it as Class II and gives the reason as "Boston Scientific is initiating this Medical Device Field Correction of the Expel" APD and APDL Drainage Catheters. To date, Boston Scientific has received one complaint for device fragmentation after an Expel" APD...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1336-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific: Expel" Drainage Catheter with Twist-Loc" Hub APDL Drainage Catheter and APDL Drainage Catheter Kit, Expel" Drainage Catheter APD Drainage Catheter, Expel" Large Capacity Drainage Catheter APD Large Capacity Drainage Catheter.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
Boston Scientific is initiating this Medical Device Field Correction of the Expel" APD and APDL Drainage Catheters. To date, Boston Scientific has received one complaint for device fragmentation after an Expel" APD catheter was implanted in the biliary system. The most serious foreseeable patient risk for this issue is additional intervention for fragment retrieval using minimally invasive meth
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Boston Scientific sent an "Urgent Medical Device Correction" letter dated February 25, 2015, to all affected customers. The letter described the problem and the product involved in the recall. The letter informed customers not to use the Expel APD and APDL Drainage catheters for bile drainage and to complete and return the Customer Acknowledgement Form. The letter also noted that the devices are not being retrieved and they are not required to return them to Boston Scientific. For further questions please call (763) 494-7971.
Quantity in commerce
918
Distribution
Worldwide Distribution - US Distribution to the states of : CA, LA, FL, WI, NY, IL, MD and OH., and to the countries of: Belgium, Czech Republic, Finland, France, Germany, Great Britain, Italy, Spain, Switzerland, Austria, Hungary, Netherlands and Portugal.
Product codes and lots
UPN # LOT #  H7493931006200- 17259582, 17362835, 17501623, 17617274. H7493931008200- 17259557, 17297984, 17362834, 17376747. H7493931010200- 17318044, 17376748, 17459140, 17501573, 17617545.  H7493931012250- 17275816, 17501620, 17575123, 17617353. H7493931014250- 17370708, 17604432, 17624928. H7493931106150- 17259581, 17365684, 17617551. H7493931108150- 17244079, 17404400, 17408751, 17493552, 17647113 H7493931108250- 17245000, 17275820, 17297985, 17297986, 17377401, 17416602, 17433865.  H749393110825K1- 17416577. H7493931108450- 17245001, 17408755, 17411332, 17493550. H7493931110150- 17317825, 17518149. H7493931110250- 17228343, 17259558, 17317546, 17338203, 17365681, 17416573, 17438002, 17518150, 17588842, 17588844, 17621339, 17634403, 17634404, 17639037.  H749393111025K1- 17328810, 17403872, 17416575, 17494061. H7493931110450- 17275823, 17416603, 17459143, 17617552. H7493931112250- 17259584, 17275818, 17338205, 17365682, 17601584, 17613485, 17639463.  H7493931112450- 17245002, 17259556, 17365683, 17416604, 17574538, 17617553.  H7493931114250- 17370706, 17588840, 17625218. H7493931114450- 17275817, 17589950, 17624789. H7493932016300- 17328801, 17352479, 17377392, 17408752, 17501571, 17575124, 17612692, 17613484, 17625883, 17626180, 17626181, 17626182.  H7493932018300- 17377395, 17480649, 17614751, 17617350, 17626183. H7493932024400- 17459142, 17492863. H7493932026400- 17328800, 17353160, 17377393, 17617470, 17624927, 17625216, 17625895.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1336-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1336-2015Class IITerminatedVoluntary: Firm initiated