510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1401-2015Class III

Vitek 2 AST-P635 REF 416 911 Gram Positive Susceptibility Card, 20 cards per container, IVD. Intended for use...

bioMerieux, Inc. / initiated 2015-02-25 / Terminated / root cause: manufacturing process

Biomerieux Inc began a recall of Vitek 2 AST-P635 REF 416 911 Gram Positive Susceptibility Card, 20 cards per container, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories... on . FDA classified it as Class III and gives the reason as "The outer carton label and package insert for the recalled product incorrectly lists Streptococcus pneumoniae as an organism for intended use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1401-2015
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
Vitek 2 AST-P635 REF 416 911 Gram Positive Susceptibility Card, 20 cards per container, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae.
Product code
LTW Susceptibility Test Cards, Antimicrobial
Reason for recall
The outer carton label and package insert for the recalled product incorrectly lists Streptococcus pneumoniae as an organism for intended use.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Biomerieux, Incs sent an Urgent Product Correction Notice on February 25, 2015, to all affected customers. The letter identified the product, the problem, and the actions to be taken by the customer. Customers were instructed to do the following: " Please confirm this letter has been distributed and reviewed by all appropriate personnel within your organization. " For the referenced test kits (VITEK 2 ASTP635, ASTP640 or ASTP641), review the claimed organisms in the Package Insert antibiotic table to confirm which organisms can be tested. " Refrain from using the referenced test kits to determine the susceptibility of Streptococcus pneumoniae. " Reference your Laboratory Standard Operating Procedures to identify card type used for specific organisms. " Please store this letter with your bioMrieux instrument documentation. Please complete the attached Acknowledgement Form and return it to bioMrieux, Inc. Customers with questions were instructed to contact their local BioMerieux representative. For questions regarding this recall call 314-731-8526.
Quantity in commerce
10409 cartons
Distribution
Worldwide Distribution to United Kingdom and Turkey. NO U.S. distribution.
Product codes and lots
lot 735352820, expiration 11AUG16; lot 735348620, expiration 30JUN16; lot 735346520, expiration 09JUN16; lot 735342320; expiration 28APR16; lot 735318020, expiration 29AUG15; lot 735315920, expiration 08AUG15; lot 735323710, expiration 25OCT15; lot 735310210, expiration 12JUN15

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1401-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1401-2015Class IIITerminatedVoluntary: Firm initiated
Z-1402-2015Class IIITerminatedVoluntary: Firm initiated
Z-1403-2015Class IIITerminatedVoluntary: Firm initiated