Z-1316-2015Class II

ExacTrac versions 6.x patient positioning systems are used to position patients during radiosurgery or...

Brainlab, Inc. / initiated 2015-02-16 / Terminated / root cause: software

Brainlab AG began a recall of ExacTrac versions 6.x patient positioning systems are used to position patients during radiosurgery or radiotherapy procedures on . FDA classified it as Class II and gives the reason as "The ExacTrac 6.x Patient Positioning System may incorrectly position the patient when using the ExacTrac Cone Beam CT (CBCT) with a TrueBeam-specific optional subvolume-CBCT". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1316-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Brainlab, Inc.
Product
ExacTrac versions 6.x patient positioning systems are used to position patients during radiosurgery or radiotherapy procedures..
Product code
IYE Accelerator, Linear, Medical
Reason for recall
The ExacTrac 6.x Patient Positioning System may incorrectly position the patient when using the ExacTrac Cone Beam CT (CBCT) with a TrueBeam-specific optional subvolume-CBCT.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
BrainLab sent an FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letters dated February 16, 2015, to all affected customers. The letters included instructions for customers to: 1) Do not use any actively re-reconstructed CBCT subvolumes from the Varian TrueBeam system with the Brainlab ExacTrac 6.x CBCT Import & Alignment Software module; 2) Import exclusively original, not modified, CBCT volumes into ExacTrac to use for patient positioning; and, 3) Continue to verify any ExacTrac CBCT-based correction using the ExacTrac X-ray verification and/or an external IGRT procedure as mandatory. Brainlab provided existing potentially affected ExacTrac v.6.x CBCT Import & Alignment Software customers (with a Varian TrueBeam system) with the product notification information. Brainlab will provide a software update with this issue solved to affected customers. Brainlab will actively contact the customer starting August 2015 to schedule the update installation. For further questions please call Customer Hotline + 49 89 99 15 68 44 or 1-800 597-5911 ( For US Customers only)
Quantity in commerce
24 systems (US); 36 systems (Foreign)
Distribution
Worldwide Distribution - USA and to the countries of : Australia, Austria, Canada Denmark, France, Germany, Hong Kong, Japan, Russia, Singapore, South Korea and United Kingdom.
Product codes and lots
ExacTrac versions 6.x including versions 6.0.0, 6.0.1, 6.0.2, 6.0.3, 6.0.4, 6.0.5; and v. 6.1.0.   Model/Catalog Numbers:   49971, CBCT IMPORT & ALIGNMENT SOFTWARE;  20836, EXACTRAC 6.0 VARIAN CBCT LICENSE;   The software may be contained in configured items:   48320, ET UPGRADE SOFTWARE 6.0 TO 6.1;  48320-99, ET UPGRADE SOFTWARE 6.0 TO 6.1 SLA;  49973B, ET DATA PREP / REVIEW SYSTEM;  49973A, ET DATA PREP / REVIEW SYSTEM;  49973, ET DATA PREP / REVIEW SYSTEM;  49998B, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;  49998A, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;  49998, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;  48304, ET 6.1 DATA PREP/REVIEW SYSTEM UPG. KIT;  48302, ET DATA PREP / REVIEW SYSTEM

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1316-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1316-2015Class IITerminatedVoluntary: Firm initiated