ToggleLoc; fastener, fixation, non degradable, soft tissue; Artificial Ligament Fixation Device #7 PE Ziploop...
Biomet, Inc. began a recall of ToggleLoc; fastener, fixation, non degradable, soft tissue; Artificial Ligament Fixation Device #7 PE Ziploop Extended Toggleloc Product Usage: The Toggleloc... on . FDA classified it as Class II and gives the reason as "The firm is recalling ToggleLocs and Ziploops manufactured in May and July of 2014 as they were not processed following a standard work process, which resulted in complaints of sterile packaging not being sealed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1363-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ToggleLoc; fastener, fixation, non degradable, soft tissue; Artificial Ligament Fixation Device #7 PE Ziploop Extended Toggleloc Product Usage: The Toggleloc System is a non-resorbable system intended to aid in arthroscopic and orthopedic reconstructive procedures requiring soft tissue fixation, due to injury or degenerative disease.
- Product code
- JDR Staple, Fixation, Bone
- Reason for recall
- The firm is recalling ToggleLocs and Ziploops manufactured in May and July of 2014 as they were not processed following a standard work process, which resulted in complaints of sterile packaging not being sealed.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Biomet sent an Urgent Medical Device Recall Notice letter dated March 4, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediate locate and discontinued use of the product and its return to Biomet. Customers may contact Field Action Specialist, Regulatory Compliance,at (574) 372-1570.
- Quantity in commerce
- 50 Units of each
- Distribution
- Worldwide Distribution: US (nationwide) Distribution in the states of:TX, GA, NM, NC and OK and the countries of: Japan, and the Netherlands.
- Product codes and lots
- PN: 904755 Lot: 019150 & PN: 909848 Lot: 729530
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1363-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1364-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.