Z-1820-2015Class II

Steriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece

Hill-Rom Inc. / initiated 2015-03-02 / Terminated

Hill-Rom, Inc. began a recall of Steriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece on . FDA classified it as Class II and gives the reason as "The metal cannula attached to the handpiece can rotate and become dislodged from the plastic handpiece". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1820-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
Steriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece.
Product code
HMX Cannula, Ophthalmic
Reason for recall
The metal cannula attached to the handpiece can rotate and become dislodged from the plastic handpiece.
FDA root cause
Use error
Action
Customers were notified via telephone and e-mail by 03/02/2015. The firm is requiring customers to complete a submitted response form and certificate of destruction for recalled devices by 03/31/2015. If recalled devices were further distributed the firm is asking customers to conduct a sub recall and forward notice of the recall. Public contact is Aspen Medical Europe Ltd, QA Department, Lynda McConaghy-QARA Manager, [email protected] , or +44 (0)1527 587716.
Quantity in commerce
66691 units
Distribution
Worldwide Distribution-US (nationwide) including the states of MI & PA, and the countries of Belgium, England, Italy, Czech Republic, Turkey, Greece, Ireland, Scotland, Spain, and Lebanon.
Product codes and lots
Product #s: 157400; 257400; 157500; 257500

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1820-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1820-2015Class IITerminatedVoluntary: Firm initiated