Quadrox-iD Adult Diffusion Membrane Oxygenator
Maquet Cardiovascular Us Sales, Llc began a recall of Quadrox-iD Adult Diffusion Membrane Oxygenator Product Usage: The Quadrox-iD Adult Diffusion Membrane Oxygenator is intended for use in an extracorporeal... on . FDA classified it as Class II and gives the reason as "A potential issue was detected that could result in blood leakage from the gas outlet port in all lots of Quadrox-iD Adult Diffusion Membrane Oxygenator with Bioline and Softline Coatings, Custom Tubing Packs and...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1689-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Quadrox-iD Adult Diffusion Membrane Oxygenator Product Usage: The Quadrox-iD Adult Diffusion Membrane Oxygenator is intended for use in an extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide from the blood and regulates the blood temperature. The oxygenator with diffusion membrane cannot be used for the administration of volatile anesthetic gases because the gas molecules are unable to pass through the diffusion membrane. Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.
- Product code
- DTZ Oxygenator, Cardiopulmonary Bypass
- Reason for recall
- A potential issue was detected that could result in blood leakage from the gas outlet port in all lots of Quadrox-iD Adult Diffusion Membrane Oxygenator with Bioline and Softline Coatings, Custom Tubing Packs and Diffusion Membrane HLS Set Advanced and HIT Set Advanced Models.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Maquet sent a "Field Safety Notice" accompanied by a Response Form to their affected customers on 3/5/2015 via FedEx. The Field Safety Notice identified the affected products/Part numbers and actions to be taken by the user. The customers were asked to complete the Field Safety Notice Response Form to acknowledge that they have received the notice. The completed notice can be faxed to 1-973-396-3607 or sent via email to [email protected]. For any questions, customers can contact their Maquet Sales Representative or Maquet Customer Service at 1-888-627-8383 (press option 2, followed by option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST.
- Quantity in commerce
- 10,199 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Israel, Italy, Japan, South Korea, Latvia, Lithuania, Luxembourg, Malta, Mexico, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia Montenegro, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey and Uruguay
- Product codes and lots
- Part numbers BO-HMOD 70000, 70105.3815, BEQ-HMOD 70000, 70105,3824
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1689-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1690-2015 | Class II | Terminated | Voluntary: Firm initiated |