HLS/HIT Set Advanced with Bioline and/or Softline Coating Console
Maquet Cardiovascular Us Sales, Llc began a recall of HLS/HIT Set Advanced with Bioline and/or Softline Coating Console Product Usage: HLS/HIT Set Advanced: The CARDIO HELP System in configuration with the HLS/HIT... on . FDA classified it as Class II and gives the reason as "A potential issue was detected that could result in blood leakage from the gas outlet port in all lots of Quadrox-iD Adult Diffusion Membrane Oxygenator with Bioline and Softline Coatings, Custom Tubing Packs and...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1690-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- HLS/HIT Set Advanced with Bioline and/or Softline Coating Console Product Usage: HLS/HIT Set Advanced: The CARDIO HELP System in configuration with the HLS/HIT Set Advanced is intended to be used within the hospital environment and outside the hospital environment (for periods up to six hours), e.g. for intra- and inter-hospital transport. Due to its system combination of integrated oxygenator and centrifugal pump, the HLS Set Advanced is suitable for both extracorporeal pulmonary support and cardiovascular support, and for simultaneous cardiovascular and pulmonary support.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- A potential issue was detected that could result in blood leakage from the gas outlet port in all lots of Quadrox-iD Adult Diffusion Membrane Oxygenator with Bioline and Softline Coatings, Custom Tubing Packs and Diffusion Membrane HLS Set Advanced and HIT Set Advanced Models.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Maquet sent a "Field Safety Notice" accompanied by a Response Form to their affected customers on 3/5/2015 via FedEx. The Field Safety Notice identified the affected products/Part numbers and actions to be taken by the user. The customers were asked to complete the Field Safety Notice Response Form to acknowledge that they have received the notice. The completed notice can be faxed to 1-973-396-3607 or sent via email to [email protected]. For any questions, customers can contact their Maquet Sales Representative or Maquet Customer Service at 1-888-627-8383 (press option 2, followed by option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST.
- Quantity in commerce
- 10, 199 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Israel, Italy, Japan, South Korea, Latvia, Lithuania, Luxembourg, Malta, Mexico, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia Montenegro, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey and Uruguay
- Product codes and lots
- Part numbers BEQ-HLS 5050, 70105,2797, BEQ-HLS 7050, 70105.2794, and BO-HLS 7050, 70105.2786
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1689-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1690-2015 | Class II | Terminated | Voluntary: Firm initiated |