Z-1357-2015Class II
RECLAIM Assembled Implant Inserter Adaptor orthopedic manual surgical instrument for hip implantation
Depuy Orthopaedics Inc. / initiated 2015-02-23 / Terminated
DePuy Orthopaedics, Inc. began a recall of RECLAIM Assembled Implant Inserter Adaptor orthopedic manual surgical instrument for hip implantation on . FDA classified it as Class II and gives the reason as "This lot of the RECLAIM Assembled Implant Inserter Adaptor Instrument may fracture during surgery and there is the potential for plastic pieces to be left in the patient". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1357-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- RECLAIM Assembled Implant Inserter Adaptor orthopedic manual surgical instrument for hip implantation.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- This lot of the RECLAIM Assembled Implant Inserter Adaptor Instrument may fracture during surgery and there is the potential for plastic pieces to be left in the patient.
- FDA root cause
- Under Investigation by firm
- Action
- The firm issued a voluntary device removal on 02/23/2015 for lot no. SO2018074 of the RECLAIM Assembled Implant Inserter Adaptor. The recall notice informs US Distributors and Medical Professionals about the issue and any actions to take. US Distributors will return affected instruments and notify one Hospital/User Facility with the Hospital Notice and Reconciliation Forms. Reconciliation forms are required for the end-user Hospital/Medical Facility Consignee. Customers may contact Mindy K. Tinsley Director Wordwide Communications Tel: (574) 372-7136 Fax: (574) 372-7717 E-mail: [email protected]
- Quantity in commerce
- 54 of Lot No: SO2018074
- Distribution
- Worldwide Distribution: US (nationwide) in CA; Ireland and Israel.
- Product codes and lots
- CATALOG NO. : 2975-00-920: Barcode (GTIN): 10603295156192 Lot No. Distribution SO2018074 (Distributed) SO2018072 (Not Distributed) SO2018073 (Not Distributed) SO2018075 (Not Distributed) SO2018080 (Not Distributed)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1357-2015 | Class II | Terminated | Voluntary: Firm initiated |