Z-1459-2015Class II
Equinoxe 44mm Short Humeral Head Catalog Number 310-01-44. Orthopedic shoulder implant component
Exactech, Inc. / initiated 2015-02-13 / Terminated / root cause: labeling and UDI
Exactech, Inc. began a recall of Equinoxe 44mm Short Humeral Head Catalog Number 310-01-44. Orthopedic shoulder implant component on . FDA classified it as Class II and gives the reason as "Incorrectly packaged. Outer and inner labeling may incorrectly identify the size of the enclosed device". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1459-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Exactech, Inc.
- Product
- Equinoxe 44mm Short Humeral Head Catalog Number 310-01-44. Orthopedic shoulder implant component.
- Product code
- KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- Reason for recall
- Incorrectly packaged. Outer and inner labeling may incorrectly identify the size of the enclosed device.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Exactech Agents sent a Product Recall Advisory Notice dated February 12, 2015 by mailed to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the Recall Inventory Respponse form and return it to Exactech, per the instructions provided on the notice.
- Quantity in commerce
- 112 devices
- Distribution
- Worldwide Distribution - US Nationwide in the states of AK, AZ, CA, CO, FL, ID, ME, MN, NC, OH, OK, PA, RI, TN, VA and WA. and countries of Brazil and Canada.
- Product codes and lots
- Serial Numbers 3597678 thru 3597713, 3609323 thru 3609358, 3616214 thru 3616249 and 3640539 thru 3640574.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1459-2015 | Class II | Terminated | Voluntary: Firm initiated |