Integra Flowable Wound Matrix Size 3cc 1 unit/box single use, sterile device Rx Only Model Number FWD301 (US)...
Integra LifeSciences Corp. began a recall of Integra Flowable Wound Matrix Size 3cc 1 unit/box single use, sterile device Rx Only Model Number FWD301 (US) & FDR301 (Non-US) Flowable Wound Matrix is... on . FDA classified it as Class II and gives the reason as "A lot of Flowable Wound Matrix that has been distributed to customers in late 2013 / early 2014 was verified as not meeting stability acceptance criteria for the attributes of visual appearance during dispensing and...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1452-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Integra Flowable Wound Matrix Size 3cc 1 unit/box single use, sterile device Rx Only Model Number FWD301 (US) & FDR301 (Non-US) Flowable Wound Matrix is comprised of the following components which are provided in a sealed and sterilized Tyvek / plastic tray: - An empty plastic syringe with Luer-Lok" tip - A plastic syringe with Luer-Lok" tip containing the Flowable Wound Matrix material - A syringe-to-syringe Luer-Lok" adapter / connector - A flexible plastic tube (injector) with Luer-Lok" connector Instructions for Use (IFU) and the sterilized package are subsequently packaged in a labeled and sealed box that is distributed to customers.
- Product code
- KGN Wound Dressing With Animal-Derived Material(S)
- Reason for recall
- A lot of Flowable Wound Matrix that has been distributed to customers in late 2013 / early 2014 was verified as not meeting stability acceptance criteria for the attributes of visual appearance during dispensing and force vs. displacement (force needed for product to be extruded from the syringe).
- FDA root cause
- Unknown/Undetermined by firm
- Action
- Integra Life Sciences Inc. sent a Recall letter/Recall acknowledgement and return form dated March 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Integra asks customers to review inventory and determine if any of the affected inventory is in possession. The attached Recall Acknowledgement and Return Form should be completed and returned. Customer service will contact the customer upon receipt of the form and provide an RMA number. Questions can be directed to Customer Service at 1-888-301-0203.
- Quantity in commerce
- 287 packages ( 179 US 108 OUS )
- Distribution
- Worldwide Distribution - US Distribution and to the countries of and to the countries of Greece, France, Italy, Spain, Switzerland, South America, Turkey and United Kingdom.
- Product codes and lots
- Model Number FWD301 (US) & FDR301 (Non-US) The following Lot Numbers on the packaged kits are affected: - 305000293250, 305000293251, 305000298616 (foreign) The syringes within the packaged kits are also Lot numbered. For the packaged kit Lot Numbers above, the corresponding respective syringe Lot Numbers are: - 10500028786, 105000288118, 105N00288965 (foreign)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1452-2015 | Class II | Terminated | Voluntary: Firm initiated |