SMARTABLATE RF Generator System, Catalog No. M490006, M490007, M4900106, and M4900107
Biosense Webster, Inc. began a recall of SMARTABLATE RF Generator System, Catalog No. M490006, M490007, M4900106, and M4900107. Product Usage: The use of the SMART ABLATE Generator and all accessories... on . FDA classified it as Class II and gives the reason as "Biosense Webster is recalling the SMARTABLATE Generator System because Biosense Webster has received two complaints reporting that during an RMT procedure the SMARTABLATE RF Generator inadvertently delivered RF energy". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1691-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biosense Webster Inc
- Product
- SMARTABLATE RF Generator System, Catalog No. M490006, M490007, M4900106, and M4900107. Product Usage: The use of the SMART ABLATE Generator and all accessories is indicated in combination with compatible therapeutic catheters for use in conventional intracardiac RF ablation procedures. The SMARTABLATE Foot Pedal is designed to initiate or discontinue operation of the SMARTABLATE System. The Foot Pedal can be connected to the SMARTABLATE RF Generator, SMARTABLATE Irrigation Pump; or SMARTABLATE Remote Control.
- Product code
- OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation
- Reason for recall
- Biosense Webster is recalling the SMARTABLATE Generator System because Biosense Webster has received two complaints reporting that during an RMT procedure the SMARTABLATE RF Generator inadvertently delivered RF energy.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Biosense Webster sent an Urgent Field Safety Notification Letter dated March 18, 2015 via FedEx to all their customers that received the affected SMARTABLATE Generator System. The letter informs the customers of the problems identified and the actions to be taken. Biosense Webster is taking precautionary safety measures of reinforcing the Warnings in the Instructions for Use (IFU) for the SMARTABLATE System Foot Pedal. Customers with questions related to the are instructed to contact their local Biosense Webster sales representative or call (866) 473-7823, Monday through Friday from 7:00AM to 8:00PM EST. Customers with questions related to the voluntary field removal certification form and its return, are instructed to contact Elie Ghattas Field Action Coordinator at (909) 839-7281.
- Quantity in commerce
- 115 units
- Distribution
- Worldwide Distribution - US Nationwide in the state of: CA, IL, KY, NJ, PA and the countries of: Czech Republic, Denmark, Finland, France, Germany, and Sweden.
- Product codes and lots
- SMARTABLATE System Serial No. G4C-0296, G4C-0297, G4C-0272, G4C-0262, G4C-0232, G4C-0090. System Serial No. G4C-0076, G4C-0195, G4C-0064, G4C-0047, G4C-0159, G4C-0187, G4C-0146, G4C-0140, G4C-0037, G4C-0131, G4C-0031, G0054.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1691-2015 | Class II | Terminated | Voluntary: Firm initiated |