Z-1369-2015Class II

ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes...

Roche Diabetes Care, Inc. / initiated 2014-10-30 / Terminated / root cause: software

Roche Diabetes Care, Inc. began a recall of ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management on . FDA classified it as Class II and gives the reason as "Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1369-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diabetes Care, Inc.
Product
ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carb
Root cause tag
software
FDA root cause
Software design
Action
Urgent Field Safety Notices were sent to the customer and healthcare professional starting on 10/30/2014. The letter provided a description of the reason for the notification; and recommended that all users should update their app immediately.
Quantity in commerce
The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
Distribution
There are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa. The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
Product codes and lots
N/A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1369-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1369-2015Class IITerminatedVoluntary: Firm initiated