Z-1392-2015Class II

da Vinci Si Surgical System IS3000, Disposable Accessory Kit, 3 ARM, 5 Pack. Microtek Medical Equipment...

Intuitive Surgical, Inc. / initiated 2015-03-16 / Terminated / root cause: packaging

Intuitive Surgical, Inc. began a recall of da Vinci Si Surgical System IS3000, Disposable Accessory Kit, 3 ARM, 5 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical... on . FDA classified it as Class II and gives the reason as "Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to tear". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1392-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
da Vinci Si Surgical System IS3000, Disposable Accessory Kit, 3 ARM, 5 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical and non-surgical equipment in various settings throughout the clinical setting.
Product code
KKX Drape, Surgical, Antimicrobial
Reason for recall
Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to tear.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Firm issued field safety notice to all customers on March 16, 2015. Field safety notice includes acknowledgement form to be returned to recalling firm. A second letter was sent on May 12, 2015 to inform users that the date that the drape product unaffected by the issue will not be availalble untisl October/November 2015, not May 2015 as stated in the original communication.
Quantity in commerce
442,475 total for all drape models
Distribution
Worldwide Distribution-Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China (including Hong Kong), Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, France, Germany, Greece, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Mauritius, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, Vietnam.
Product codes and lots
Product number 420290-03; da Vinci Si Disposable Accessory Kit, 3 ARM, 5 Pack.  Lot Numbers for Cloudy/Waxy issue: ALL lot numbers   Lot Numbers for Tear issue: D142175, D142175A, D142185, D142185A, D142235, D142235A, D142245, D142245A, D142255, D142305, D142305A, D142315, D142315A, D142345, D142375, D142375A, D142385, D142385A, D142395, D142455, D142465, D142465A, D142475, D142475A, D142485, D142515, D142515A, D142525, D142525A, D142535, D142535A, D142605, D142605A, D142615, D142615A, D142625, D142635, D142655, D142655A, D142665, D142665A, D142725, D142725A, D142735, D142735A, D142745, D142755, D142755A, D142795, D142795A, D142825, D142825A, D142835, D142885, D142885A, D142885A, D142895, D142905, D142945, D142955, D142955A, D142965, D142965A, D143005, D143015, D143015A, D143025, D143025A, D143035, D143045, D143045A, D143095, D143095A, D143105, D143105A, D143155, D143155A, D143165, D143165A, D143175, D143175A, D143185, D143185A, D143215, D143215A, D143225, D143225A, D143235, D143235A, D143295, D143295A, D143305, D143305A, D143315, D143325, D143325A, D143365, D143375, D143375A, D143385, D143395, D143395A, D143435, D143445, D143445A, D143455, D143455A, D143465, D143465A, D150075, D150085, D150125, D150125A, D150135, D150165, D150195, D150195A, D150205, D150205A, D150275, D150275A, D150285, D150305, D150345, D150345A, D150355, D150355A, D150415, D150415A, D150425, D150425A, D150435, D150475, D150475A, D150485, D150485A, D150495, D150545, D150545A, D150555, D150555A, D150635, D150635A, D150645, D150645A, D150685, D150685A, D150695, D150695A, DA142235, DA142305, DA142515, DA142615, DA142795, DA142885, DA142885, DA143025, DA150125, DA150275, DA150475, DA150545, DA150645, DA150685

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1392-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1380-2015Class IITerminatedVoluntary: Firm initiated
Z-1381-2015Class IITerminatedVoluntary: Firm initiated
Z-1382-2015Class IITerminatedVoluntary: Firm initiated
Z-1383-2015Class IITerminatedVoluntary: Firm initiated
Z-1384-2015Class IITerminatedVoluntary: Firm initiated
Z-1385-2015Class IITerminatedVoluntary: Firm initiated
Z-1386-2015Class IITerminatedVoluntary: Firm initiated
Z-1387-2015Class IITerminatedVoluntary: Firm initiated
Z-1388-2015Class IITerminatedVoluntary: Firm initiated
Z-1389-2015Class IITerminatedVoluntary: Firm initiated
Z-1390-2015Class IITerminatedVoluntary: Firm initiated
Z-1391-2015Class IITerminatedVoluntary: Firm initiated
Z-1392-2015Class IITerminatedVoluntary: Firm initiated
Z-1393-2015Class IITerminatedVoluntary: Firm initiated